A flowerpot-shaped birdcage coil designed for breast MRI at 0.35T
Motivation: Low-field MRI breast screening is important for the well-being of women. However, the vertical orientation of the B0 field restricts the coil design. Goal(s): To design a breast coil for low-field MR system that improves the signal-to-noise ratio(SNR) and receiving sensitivity(RS) while conforming to the anatomical features of breast. Approach: We designed an improved birdcage structure for breast imaging. It adapts to the breast shape and enhances SNR. Additionally, we used metamaterial to further improve image quality. Results: The average SNR of flowerpot coil is 162, which is approximately 1.45 times higher than the commercial coil. The Q-ratio, approximately 1.30, indicates good RS. Impact: We proposed a novel flowerpot-shaped birdcage coil for breast MRI using metamaterials in a 0.35T system, resulting in an improved signal-to-noise ratio (SNR) and receiving sensitivity(RS).
Authors
- Zhenan Jiang (ORCID: https://orcid.org/0000-0002-3482-3510)
- Bensheng Qiu (ORCID: https://orcid.org/0000-0003-2987-7378)
- Lin Chen (ORCID: https://orcid.org/0000-0003-4798-6199)
- Qingyun Liu (ORCID: https://orcid.org/0000-0003-4815-3463)
- Xueyan Song
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/1642
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00