Pediatric Liver MRI: A Comparison of Image Quality and Scanning Efficiency between StarVIBE Dixon and VIBE Dixon Sequences
Motivation: To addresses the challenge of low breath-holding cooperation in pediatric abdominal MRI examinations. Goal(s): To compare the image quality and scanning efficiency of StarVIBE Dixon and VIBE Dixon in pediatric liver. Approach: Fifty pediatric patients underwent abdominal MRI examinations using the StarVIBE Dixon and VIBE Dixon sequences, with subjective and objective evaluation and comparative analysis of image quality. Results: Both sets of images met the clinical diagnostic criteria. The image quality of StarVIBE Dixon was slightly superior to that of VIBE Dixon (P<0.05), , VIBE Dixon is more appropriate for scans performed after sedation to decrease the risk of anesthesia exposure. Impact: This study identifies the optimal MRI sequences for pediatric liver scans in different clinical scenarios, aiming to enhance diagnostic accuracy and reduce patient discomfort and anesthesia-related risks.
Authors
- 鹏 张
- 政远 李
- 跃文 郝
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/5297
- Primary Topic
- Liver Disease Diagnosis and Treatment
- Type
- article
- Field-Weighted Citation Impact
- 0.00