Gradient swept adiabatic pulses for out of slice inversion and outer volume suppression.
Motivation: Slice selective adiabatic pre-pulses have many uses in magnetic resonance. Using gradient waveforms to create adiabatic rotations creates new ways to prepare magnetization. Goal(s): Explore adiabatic pulse with frequency sweeps created by gradient wave forms instead of RF frequency waveforms. Approach: Gradient swept adiabatic inversion pulses were tested with numerical Bloch simulations and on a phantom in a 3T scanner. Results: The pulses can invert magnetization outside a well-defined non-perturbed slice. Sub-adiabatic use allows outer volume suppression, and the non-perturbed slice can be transformed into a wedge with additional constant orthogonal gradients. Impact: A new class of adiabatic inversion pulses with frequency sweeps generated by gradient waveforms to yield inversion outside a well-defined non-perturbed slice or wedge delivers new options for pre-pulses for contrasts, spatial selection, or outer volume suppression.
Authors
- Ahmed M. Gharib (ORCID: https://orcid.org/0000-0002-2476-481X)
- Ronald Ouwerkerk (ORCID: https://orcid.org/0000-0003-3882-9757)
- Nader S. Metwalli
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4422
- Primary Topic
- Laser-Matter Interactions and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00