Simulation of SAR and B1+ homogeneity change with subject movement using 16ch transceiver coil at 7T
Motivation: SAR estimation at UHF is critical. Particularly when subject moves during scan, SAR could be changed significantly, potentially exceeding limit. Goal(s): We investigated how SAR and B1+ homogeneity is changed by head movement, coil tunning/matching(TM) and B1+ shim. Approach: Using EM simulation with human model with 16ch transceiver, B1+ and SAR distribution was estimated for different head positions, coil TM and B1+ shim. Results: Head movement is main contributor to SAR increase; but also suboptimal coil TM and B1+ shim contributing to the variation. Impact: Estimation of B1+ and SAR changes by subject motion would improve scan safety as well as B1+ shim approach.
Authors
- Donghyuk Kim (ORCID: https://orcid.org/0000-0003-1381-3943)
- Kyoung‐Nam Kim (ORCID: https://orcid.org/0000-0001-7502-2851)
- Chan Hong Moon
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/3965
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00