Cardiac T1-mapping using Open-MOLLI at 3T: a repeatability study
Motivation: Myocardial T1-mapping was introduced to improve repeatability compared to qualitative measurements but extracted T1 values vary across scanners. The use of vendor-neutral techniques provides a robust and accessible approach for standardizing measurements, but their repeatability still needs to be assessed. Goal(s): In this work we aim to study the repeatability of open-source myocardial T1 mapping (Open-MOLLI) at 3.0T and compare it to the vendor-implemented MOLLI sequence. Approach: MOLLI and Open-MOLLI sequences were acquired in 15 subjects with repetition in each session to study T1 repeatability. Results: Similar repeatability for in vivo T1 values was observed for MOLLI and Open-MOLLI at 3.0T. Impact: The open-source myocardial T1 mapping (Open-MOLLI) was tested at 3.0T for repeatability with similar results to previous works at 1.5 T. This work shows that this sequence can be easily applied in inter-scanner reproducibility studies.
Authors
- Andreia Gaspar (ORCID: https://orcid.org/0000-0003-2044-6213)
- Thomas Küstner (ORCID: https://orcid.org/0000-0002-0353-4898)
- Martin Schwartz (ORCID: https://orcid.org/0000-0002-9006-7844)
- Teresa Correia (ORCID: https://orcid.org/0000-0002-1606-9550)
- Rita G. Nunes (ORCID: https://orcid.org/0000-0001-7425-5717)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/1665
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00