An Automated Validation System for Active Implantable Medical Device (AIMD) Model under MR RF Field Exposure
Motivation: Current AIMD model validation method is expensive, time-consuming, and subject to human operational error. An automated robotic system for testing would streamline validation testing. Goal(s): To develop a novel automated AIMD model validation system that is cost-effective and accurate. Approach: A novel system is developed by integrating the a robotic arm, a box-shaped phantom, a signal generator and power amplifier, and a fiber optic temperature measurement apparatus to streamline the validation. Results: An automated validation system for both 1.5T and 3T is constructed with minimum human error. Impact: A cost-effective and yet more accurate AIMD model validation system/method is developed following ISO 10974 standard. Compared to the traditional birdcage coil validation method, this system requires less labor, lower cost, and has lower systematic error.
Authors
- Hongbae Jeong (ORCID: https://orcid.org/0000-0003-4908-2070)
- Farshad Ebrahimi (ORCID: https://orcid.org/0009-0009-5798-3160)
- Qingyan Wang (ORCID: https://orcid.org/0000-0001-6104-4048)
- Ao Shen
- Ananda Kumar
- Ziyu Zuo
- Ji Chen
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4978
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00