Non-contrast measurement of BBB permeability using an accelerated WEPCAST MRI with velocity-selective single-shot acquisition

Motivation: WEPCAST MRI was proposed for non-contrast measurement of BBB permeability, but is limited to long scan duration and sensitive to motion due to its double-subtraction scheme. Goal(s): This study aimed to design a motion-robust and accelerated WEPCAST sequence. Approach: The proposed sequence employed a velocity-selective single-shot acquisition as an alternative to the original phase-contrast for isolating the arterially-labeled spins in the vein. Sequence parameters were optimized. The validity and reproducibility were also examined. Results: VS-WEPCAST reduced scan duration from 6 to 3 minutes, while keeping a strong consistency with the original WEPCAST technique. Test-retest results indicate good reproducibility. Impact: This study proposed an accelerated velocity-selective WEPCAST technique that enables the assessment of global BBB permeability to water in nearly half the original scan time, with improved robustness to motion and cardiac pulsation.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/0728
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Non-contrast measurement of BBB permeability using an accelerated WEPCAST MRI with velocity-selective single-shot acquisition

Yung‐Chin Hsu, Zhao Li, Dan Wu, Zixuan Lin et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Non-contrast measurement of BBB permeability using an accelerated WEPCAST MRI with velocity-selective single-shot acquisition

Yung‐Chin Hsu, Zhao Li, Dan Wu, Zixuan Lin, Yizhe Hu, Hanzhang Lu
article en

Abstract

Motivation: WEPCAST MRI was proposed for non-contrast measurement of BBB permeability, but is limited to long scan duration and sensitive to motion due to its double-subtraction scheme. Goal(s): This study aimed to design a motion-robust and accelerated WEPCAST sequence. Approach: The proposed sequence employed a velocity-selective single-shot acquisition as an alternative to the original phase-contrast for isolating the arterially-labeled spins in the vein. Sequence parameters were optimized. The validity and reproducibility were also examined. Results: VS-WEPCAST reduced scan duration from 6 to 3 minutes, while keeping a strong consistency with the original WEPCAST technique. Test-retest results indicate good reproducibility. Impact: This study proposed an accelerated velocity-selective WEPCAST technique that enables the assessment of global BBB permeability to water in nearly half the original scan time, with improved robustness to motion and cardiac pulsation.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 65%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.