High SNR Efficiency Diffusion-Prepared Imaging with Amplitude Navigation
Motivation: To improve the SNR efficiency of diffusion-prepared imaging by avoiding the use of a stabilizer gradient and its 2× signal penalty. Goal(s): Demonstrate that shot-to-shot motion-induced phases can be modelled as image-based amplitude variations in reconstruction. Approach: Monte Carlo simulations showed that eliminating the stabilizer gradient increases the number of excitations required to obtain accurate ADC estimates. Diffusion-prepared acquisitions with and without stabilizer gradients were performed in the brain and reconstructed with different retrospective NEX. Results: Eliminating the stabilizer gradient improved SNR at the cost of an increased minimum scan time, and diffusion contrast sensitivity to B1 inhomogeneity. Impact: Replacing the stabilizer gradient with use of a signal amplitude correction in diffusion-prepared imaging can reliably improve the SNR efficiency in cases where motion-induced and eddy-current-induced phase is controlled.
Authors
- Philip Lee (ORCID: https://orcid.org/0000-0002-5783-1599)
- Xuetong Zhou (ORCID: https://orcid.org/0009-0007-2584-0558)
- Brian Hargreaves
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4191
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00