Application of an IEC 62464-1 MR Image Quality System for Standard Quality Assurance in Fat-Suppressed MRI near Metallic Implants
Motivation: The researchers are motivated by implementation of regular QA MRI in accordance with the international IEC 62464-1 guidelines and its applications in fat-suppressed MRI near metals. Goal(s): To evaluate the potential of IEC 62461-1 standard for regular QA scanning in fat suppressed MRI near metals. Approach: A designated phantom developed according to IEC 62464-1 was evaluated in regular QA MRI. Data quality was assessed in presence of metals along with fat suppression. Results: The results demonstrate QA MRI stability of the investigated system over time. Application of the QA phantom enables quantification of metallic implant influence and fat suppression on MRI image quality. Impact: The use of the IEC standard is expected to deliver a generalised approach ensuring the best data quality in fat-suppressed Gd-based MRI in presence of metals.
Authors
- Andreas Melzer (ORCID: https://orcid.org/0000-0002-7874-8946)
- Denis M. Kokorin (ORCID: https://orcid.org/0000-0002-4842-7711)
- Gregor Schaefers
- Leon Schuelert
- Andreas Eger
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/5428
- Primary Topic
- Advanced X-ray and CT Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00