A Low Inductance Nonlinear Gradient Array Design to Mitigate PNS and Reduce Voltage for Extension to Ultrasonic Whole-Body Fast and Silent MRI
Motivation: A nonlinear gradient has the potential to facilitate whole-body ultrasonic silent encoding at 20 kHz while avoiding PNS. Yet, high voltage concerns at 20 kHz switching limits the maximum number of windings possible in a nonlinear gradient design. Goal(s): To demonstrate encoding with a nonlinear gradient field produced by a gradient-array design. Approach: We designed a prototype nonlinear gradient-array and imaged a phantom at 20 kHz. Data was reconstructed using a PSF-based method. Results: The gradient-array enabled a nonlinear gradient shape to be generated with lower inductance than is possible with a single series circuit, reducing the induced voltage. Impact: Whole-body silent MRI is necessary to improve patient comfort. Here, we show a nonlinear gradient-array is a viable option to avoid PNS while achieving ultrasonic switching rates. This suggests whole-body fast and silent MRI is possible with a gradient-array design.
Authors
- Jeroen C.W. Siero (ORCID: https://orcid.org/0000-0001-5079-2868)
- Dennis W. J. Klomp (ORCID: https://orcid.org/0000-0002-5884-5386)
- K. S. Meng
- Thomas Roos (ORCID: https://orcid.org/0000-0002-3465-7794)
- Edwin Versteeg (ORCID: https://orcid.org/0000-0003-3235-3970)
- Michael McGrory (ORCID: https://orcid.org/0009-0008-1364-6595)
- Erik van Riel
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/1337
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00