Simultaneous DCE-DSC evaluation with 3D stack-of-spiral FLASH readout

Motivation: Dynamic contrast enhanced (DCE) and dynamic susceptibility contrast (DSC) are both essential for assessing cerebrovascular function, but their accuracy is often comprised by competing relaxation effects. Goal(s): We aimed to develop a straightforward method enabling simultaneous DCE and DSC evaluation to improve the accuracy of assessment. Approach: Dynamic T1 and T2* mapping was simultaneously achieved using a dual flip angle with multi-echo (DFA-ME) sequence based on efficient 3D stack-of-spiral FLASH readout. Results: DFA-ME demonstrated strong correlation with the reference method for T1 and T2* mapping. Additionally, the permeability and perfusion parameters derived from DFA-ME were consistent with values reported in the previous literature. Impact: DFA-ME enables whole-brain dynamic T1 and T2* mapping at 2.0×2.0×6.0 mm3 spatial resolution and a temporal resolution of 2.1 s, allowing simultaneous and accurate evaluation of DCE and DSC with a single-dose contrast injection.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/4778
Primary Topic
Nuclear Physics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simultaneous DCE-DSC evaluation with 3D stack-of-spiral FLASH readout

Soonyoung Yu, Xuejing Lin, Huijun Chen, Xiaoming Liu et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Nuclear Physics and Applications
article

Simultaneous DCE-DSC evaluation with 3D stack-of-spiral FLASH readout

Soonyoung Yu, Xuejing Lin, Huijun Chen, Xiaoming Liu, Rui Li, Zhiyong Xu, Hancong Sun
article en

Abstract

Motivation: Dynamic contrast enhanced (DCE) and dynamic susceptibility contrast (DSC) are both essential for assessing cerebrovascular function, but their accuracy is often comprised by competing relaxation effects. Goal(s): We aimed to develop a straightforward method enabling simultaneous DCE and DSC evaluation to improve the accuracy of assessment. Approach: Dynamic T1 and T2* mapping was simultaneously achieved using a dual flip angle with multi-echo (DFA-ME) sequence based on efficient 3D stack-of-spiral FLASH readout. Results: DFA-ME demonstrated strong correlation with the reference method for T1 and T2* mapping. Additionally, the permeability and perfusion parameters derived from DFA-ME were consistent with values reported in the previous literature. Impact: DFA-ME enables whole-brain dynamic T1 and T2* mapping at 2.0×2.0×6.0 mm3 spatial resolution and a temporal resolution of 2.1 s, allowing simultaneous and accurate evaluation of DCE and DSC with a single-dose contrast injection.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 69%
Nuclear Physics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.