Robust T2 Mapping in the Presence of Large B1+ Inhomogeneities
Motivation: Large B1+ variations near metallic implants results in errors in quantitative measurements. Goal(s): We propose a robust T2 relaxometry technique that accounts for large B1+ inhomogeneities near metallic implants. Approach: We developed a multi-echo TSE sequence that uses variable excite and refocusing flip angles to sensitize the sequence to both T2 and B1+ variations. We propose the use of two fitting algorithms to estimate B1+ robust T2 estimates. Results: In the presence of metallic implants, the proposed technique allows improved estimation of the B1+ variations which in turn improves the T2 estimation. Impact: The use of a B1+ sensitized TSE sequence to estimate T2 relaxometry can improve estimation performance in areas of large B1+ inhomogeneities near metallic implants. This would allow quantitative assessment and characterization of bone pathology near the arthroplastic implant interface.
Authors
- Jan Fritz (ORCID: https://orcid.org/0000-0003-4456-3043)
- Iman Khodarahmi (ORCID: https://orcid.org/0000-0003-3067-2345)
- Mahesh Keerthivasan (ORCID: https://orcid.org/0000-0002-7841-9333)
- Mary T. Bruno (ORCID: https://orcid.org/0000-0003-1625-0293)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/4965
- Primary Topic
- Characterization and Applications of Magnetic Nanoparticles
- Type
- article
- Field-Weighted Citation Impact
- 0.00