Enhancing Reliability of MRI-based Brain Morphometry by Synthetic MPRAGE Generation

Motivation: Brain morphometry is increasingly recognized as potential biomarker for tracking neurodegenerative disease progression. However, variations in MRI acquisition parameters, common in clinical practice, compromise the reliability of morphometric measures. Goal(s): To achieve contrast-invariant morphometry across MPRAGE images with varying grey/white matter contrasts. Approach: We retrained a deep-learning-based segmentation tool (DL+DiReCT) by modeling MPRAGE contrast variations as a function of MRI parameters (TI and TR). Cortical thickness variability was assessed before and after retraining. Results: The retrained model improved contrast invariance, reducing the contrast-related cortical thickness dependence from 5% to 2% across clinically relevant parameters. Impact: We created a contrast-invariant segmentation tool that improves brain morphometry accuracy across variable MRI settings, enabling more reliable monitoring of neurodegenerative disease progression. This tool improves assessment accuracy across longitudinal and multi-parameter MRI acquisitions common in clinical practice.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/4314
Primary Topic
Medical Image Segmentation Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhancing Reliability of MRI-based Brain Morphometry by Synthetic MPRAGE Generation

Christian Rummel, Richard McKinley, Milena Capiglioni, Roland Wiest et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Medical Image Segmentation Techniques
article

Enhancing Reliability of MRI-based Brain Morphometry by Synthetic MPRAGE Generation

Christian Rummel, Richard McKinley, Milena Capiglioni, Roland Wiest, Timo Blattner
article en

Abstract

Motivation: Brain morphometry is increasingly recognized as potential biomarker for tracking neurodegenerative disease progression. However, variations in MRI acquisition parameters, common in clinical practice, compromise the reliability of morphometric measures. Goal(s): To achieve contrast-invariant morphometry across MPRAGE images with varying grey/white matter contrasts. Approach: We retrained a deep-learning-based segmentation tool (DL+DiReCT) by modeling MPRAGE contrast variations as a function of MRI parameters (TI and TR). Cortical thickness variability was assessed before and after retraining. Results: The retrained model improved contrast invariance, reducing the contrast-related cortical thickness dependence from 5% to 2% across clinically relevant parameters. Impact: We created a contrast-invariant segmentation tool that improves brain morphometry accuracy across variable MRI settings, enabling more reliable monitoring of neurodegenerative disease progression. This tool improves assessment accuracy across longitudinal and multi-parameter MRI acquisitions common in clinical practice.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 76%
Medical Image Segmentation Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Enhancing Reliability of MRI-based Brain Morphometry by Synthetic MPRAGE Generation — Christian Rummel, Richard McKinley, et al. · Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition (2025) | TGRS Research Map | TGRS