Design of Asymmetric Permanent Magnet Arrays for the development of a Portable MR Head Imager
Motivation: To propose a new permanent magnet array with high field homogeneity and compatible with existing MRI technologies. Goal(s): Our goal was to design an asymmetric permanent magnet array to significantly improve the field homogeneity and make it easier to manufacture. Approach: The proposed permanent magnet array incorporates an asymmetric arrangement of magnet elements, combining elliptical and circular arrays. Additionally, shim magnets and multi-coil are employed to improve the homogeneity. Results: The proposed array produces a field with an average strength of 55 mT and an inhomogeneity of 16984 ppm within 20 cm DSV. After shimming, the inhomogeneity was reduced to 3665 ppm. Impact: The proposed asymmetric permanent magnet array significantly improves homogeneity, making it a promising option for portable MRI brain imaging system.
Authors
- Zhihua Ren (ORCID: https://orcid.org/0000-0003-2249-6484)
- Zongchao Shao
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/0904
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00