Design of Asymmetric Permanent Magnet Arrays for the development of a Portable MR Head Imager

Motivation: To propose a new permanent magnet array with high field homogeneity and compatible with existing MRI technologies. Goal(s): Our goal was to design an asymmetric permanent magnet array to significantly improve the field homogeneity and make it easier to manufacture. Approach: The proposed permanent magnet array incorporates an asymmetric arrangement of magnet elements, combining elliptical and circular arrays. Additionally, shim magnets and multi-coil are employed to improve the homogeneity. Results: The proposed array produces a field with an average strength of 55 mT and an inhomogeneity of 16984 ppm within 20 cm DSV. After shimming, the inhomogeneity was reduced to 3665 ppm. Impact: The proposed asymmetric permanent magnet array significantly improves homogeneity, making it a promising option for portable MRI brain imaging system.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/0904
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design of Asymmetric Permanent Magnet Arrays for the development of a Portable MR Head Imager

Zhihua Ren, Zongchao Shao
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Design of Asymmetric Permanent Magnet Arrays for the development of a Portable MR Head Imager

Zhihua Ren, Zongchao Shao
article en

Abstract

Motivation: To propose a new permanent magnet array with high field homogeneity and compatible with existing MRI technologies. Goal(s): Our goal was to design an asymmetric permanent magnet array to significantly improve the field homogeneity and make it easier to manufacture. Approach: The proposed permanent magnet array incorporates an asymmetric arrangement of magnet elements, combining elliptical and circular arrays. Additionally, shim magnets and multi-coil are employed to improve the homogeneity. Results: The proposed array produces a field with an average strength of 55 mT and an inhomogeneity of 16984 ppm within 20 cm DSV. After shimming, the inhomogeneity was reduced to 3665 ppm. Impact: The proposed asymmetric permanent magnet array significantly improves homogeneity, making it a promising option for portable MRI brain imaging system.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 65%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Design of Asymmetric Permanent Magnet Arrays for the development of a Portable MR Head Imager — Zhihua Ren, Zongchao Shao · Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition (2025) | TGRS Research Map | TGRS