Toward Robust Fat-Suppressed Gadolinium-Enhanced MRI Near Metal Implants: A Local B₀ Shim Coil Insert Approach
Motivation: Conventional fat suppression often fails near metallic implants due to severe B₀ inhomogeneity. This failure can obscure gadolinium-enhanced lesions. Goal(s): A method for robust fat suppression during gadolinium-enhanced MRI near metal implants. Approach: A 7-channel local B₀ shim coil insert was developed. Phantom and in vivo experiments were performed on a 3 T MRI scanner, comparing the shim insert to standard second-order SH shimming. Results: With ~37% reduced B₀ RMSE, the shim insert suppressed fat signals that remained bright under standard shimming, enabling reliable fat-suppressed MRI despite severe B₀ inhomogeneity. Impact: This hardware-based local B₀ shim approach can transform musculoskeletal MRI near metal implants by enabling uniform fat suppression and clarifying gadolinium-enhanced lesions, thus improving clinical diagnoses. Future investigations may expand this strategy to broader anatomies and more severe off-resonance scenarios.
Authors
- Nader Binesh
- Hui Han (ORCID: https://orcid.org/0000-0003-4440-4421)
- Fardad Serry
- Jeremy Zepeda (ORCID: https://orcid.org/0000-0001-8955-2992)
- Yun Shang (ORCID: https://orcid.org/0000-0003-3429-7730)
- Ziyang Long
- Archana Malagi (ORCID: https://orcid.org/0000-0002-2332-170X)
- Debiao Li (ORCID: https://orcid.org/0000-0001-8560-8231)
- Li-Ting Huang
- Hsin-Jung Yang
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/5426
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00