Reducing aliasing-induced artifacts in BLADE sequence using extended-FOV reconstruction
Motivation: BLADE (PROPELLER) is robust to motion but more prone to aliasing-induced artifacts. Goal(s): To minimize aliasing-induced artifacts in BLADE(PROPELLER) sequence. Approach: Utilizing GRAPPA method to extend FOV for each blade in reconstruction. Pre-scan data is used to calibrate the GRAPPA kernels for each blade. Results: Proposed method can effectively eliminate aliasing-induced artifacts. Impact: The extended-FOV method can be integrated seamlessly into the existing scanning process without requiring additional scan time, to effectively eliminate aliasing-induced artifacts.
Authors
- Kun Zhou (ORCID: https://orcid.org/0000-0003-0650-9521)
- Nan Xiao (ORCID: https://orcid.org/0009-0003-0886-9982)
- Yang Li (ORCID: https://orcid.org/0000-0002-9544-5840)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/5134
- Primary Topic
- Algorithms and Data Compression
- Type
- article
- Field-Weighted Citation Impact
- 0.00