Comprehensive Assessment of Nonuniform Image Quality: Replication and Application to Metal Imaging at 0.55T

Motivation: To evaluate nonuniform image quality in 0.55T MRI. Goal(s): To replicate a recently published method for comprehensive assessment of nonuniform image quality1 in the context of imaging around metal at 0.55T. Approach: A total hip arthroplasty (THA) implant and plastic replica were scanned using 2D-TSE and SEMAC. Intensity artifact, SNR, geometric distortion and spatial resolution maps were estimated. Results: We successfully replicated a method for comprehensive assessment of nonuniform image quality, and demonstrate its ability to quantify intensity artifacts, geometric distortion, signal-to-noise ratio, and spatial resolution. Impact: We demonstrate successful replication of the original methods paper, utilizing conventional 3D printing, a gyroid lattice phantom, and a THA phantom with a cobalt-chromium head, for the application to MRI near metallic implants at 0.55T.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/2588
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comprehensive Assessment of Nonuniform Image Quality: Replication and Application to Metal Imaging at 0.55T

Alexander Toews, Bahadır Alp Barlas, Krishna S. Nayak, Ana Sanson Leon et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced X-ray and CT Imaging
article

Comprehensive Assessment of Nonuniform Image Quality: Replication and Application to Metal Imaging at 0.55T

Alexander Toews, Bahadır Alp Barlas, Krishna S. Nayak, Ana Sanson Leon, Brian Hargreaves
article en

Abstract

Motivation: To evaluate nonuniform image quality in 0.55T MRI. Goal(s): To replicate a recently published method for comprehensive assessment of nonuniform image quality1 in the context of imaging around metal at 0.55T. Approach: A total hip arthroplasty (THA) implant and plastic replica were scanned using 2D-TSE and SEMAC. Intensity artifact, SNR, geometric distortion and spatial resolution maps were estimated. Results: We successfully replicated a method for comprehensive assessment of nonuniform image quality, and demonstrate its ability to quantify intensity artifacts, geometric distortion, signal-to-noise ratio, and spatial resolution. Impact: We demonstrate successful replication of the original methods paper, utilizing conventional 3D printing, a gyroid lattice phantom, and a THA phantom with a cobalt-chromium head, for the application to MRI near metallic implants at 0.55T.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 76%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.