Comprehensive Assessment of Nonuniform Image Quality: Replication and Application to Metal Imaging at 0.55T
Motivation: To evaluate nonuniform image quality in 0.55T MRI. Goal(s): To replicate a recently published method for comprehensive assessment of nonuniform image quality1 in the context of imaging around metal at 0.55T. Approach: A total hip arthroplasty (THA) implant and plastic replica were scanned using 2D-TSE and SEMAC. Intensity artifact, SNR, geometric distortion and spatial resolution maps were estimated. Results: We successfully replicated a method for comprehensive assessment of nonuniform image quality, and demonstrate its ability to quantify intensity artifacts, geometric distortion, signal-to-noise ratio, and spatial resolution. Impact: We demonstrate successful replication of the original methods paper, utilizing conventional 3D printing, a gyroid lattice phantom, and a THA phantom with a cobalt-chromium head, for the application to MRI near metallic implants at 0.55T.
Authors
- Alexander Toews (ORCID: https://orcid.org/0009-0007-6682-1708)
- Bahadır Alp Barlas (ORCID: https://orcid.org/0000-0001-6963-9644)
- Krishna S. Nayak (ORCID: https://orcid.org/0000-0001-5735-3550)
- Ana Sanson Leon
- Brian Hargreaves
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/2588
- Primary Topic
- Advanced X-ray and CT Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00