Optimization of Lipid Inversion with Hyperbolic Secant Adiabatic Pulses in Transient NOE at 7T
Motivation: In vivo lipid quantification in brain tissue is challenging due to limited uniform spin inversion with conventional adiabatic pulses, affecting transient NOE (tNOE) reliability. Goal(s): Optimize B1 amplitude of hyperbolic secant (HS) adiabatic pulses in tNOE to reduce magnetization transfer (MT) effects and improve lipid quantification repeatability. Approach: Validated optimized pulse parameters on a water phantom, then acquired repeatable tNOE maps in vivo, assessing white and gray matter contrast. Results: Optimized scanner voltage (305 V) produced high-quality tNOE contrast with minimal MT interference, showing CoVs of 1% for white and 3% for gray matter. Impact: This approach may improve lipid degeneration monitoring in neurodegenerative diseases like MS and Alzheimer's.
Authors
- Ravinder Reddy (ORCID: https://orcid.org/0000-0003-4580-2392)
- Blake Benyard (ORCID: https://orcid.org/0000-0002-2480-4164)
- Dushyant Kumar (ORCID: https://orcid.org/0000-0002-1617-6399)
- Neil Wilson (ORCID: https://orcid.org/0009-0002-7766-2929)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/1498
- Primary Topic
- Spectroscopy Techniques in Biomedical and Chemical Research
- Type
- article
- Field-Weighted Citation Impact
- 0.00