Investigating the Impact of Spatially Selective Signal Suppression on BOLD fMRI Reliability
Motivation: Reduced field-of-view (rFOV) imaging can improve spatio-temporal resolution in functional MRI (fMRI), but its impact on activation detection has not (to our knowledge) been examined. Goal(s): To design a short (6ms) outer-volume-suppression (OVS) RF pulse for rFOV fMRI, and investigate its impact on the sensitivity and reliability of BOLD fMRI activation. Approach: We obtained accelerated (R=7.5) BOLD data during four repetitions of a block finger-tapping task, with and without OVS. We assessed test-retest reliability of the activation maps using receiver-operating-characteristic (ROC) analysis. Results: The 6 ms OVS pulse somewhat lowers sensitivity to motor cortex BOLD activation, but somewhat improves test-retest reliability. Impact: Outer-volume suppressed fMRI can produce reliable activation maps, and may be a promising candidate for high-resolution (e.g., layer-specific) fMRI.
Authors
- Yongli He (ORCID: https://orcid.org/0000-0002-5607-4194)
- Jon‐Fredrik Nielsen (ORCID: https://orcid.org/0000-0002-2058-3579)
- Luis Hernández-García (ORCID: https://orcid.org/0000-0003-3002-0304)
- Douglas C. Noll (ORCID: https://orcid.org/0000-0002-0983-3805)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/2334
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00