Improving Image Resolution in Deuterium Metabolic Imaging with bSSFP for Clinical Translation at 3T
Motivation: Improve SNR and spatial resolution in deuterium metabolic imaging (DMI) on 3T in humans. Goal(s): Implement and validate a bSSFP DMI sequence; investigate whether up-take alone can be a measure for DMI. Approach: The DMI bSSFP sequence excites the whole volume with a block RF pulse, followed by 3D phase encoding, acquisition and phase rewinder. Three spatial dimensions are density-weighted phase-encoded, while spectral encoding is insufficient for separating different metabolites but used for artefact reduction. Results: Voxel size in human brain was improved from (2.4cm)3 for regular, spectrally-resolved MRSI to (1.5cm)3 for bSSFP MRSI in 10minutes scan time with minor loss in SNR. Impact: DMI provides valuable clinical information, such as tumour-treatment response, and might complement or even partially substitute PET. The significant gain in spatial resolution demonstrated here could be the missing link to establish DMI clinically on 3T.
Authors
- Rolf F. Schulte (ORCID: https://orcid.org/0000-0002-1334-1264)
- Esben Søvsø Szocska Hansen (ORCID: https://orcid.org/0000-0001-5512-9870)
- Christoffer Laustsen (ORCID: https://orcid.org/0000-0002-0317-2911)
- Michael Væggemose (ORCID: https://orcid.org/0000-0001-7666-6819)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/2898
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00