Impact of Combined Anterior and Posterior Surface Coils Compared to Anterior Coil and Spine Coil on SNR in Abdominal MRI at 0.55T Free Max Scanner
Motivation: Combining two contour coils in a mid-field (0.55T) MRI to enhance the (SNR) in MRI images compare to anterior surface coil arrays paired with spine coils. Goal(s): Dedicated anterior and posterior surface coils at 0.55T is enhancing abdominal MRI image quality compared to anterior coil with spine coil. Approach: This research utilizes both anterior and posterior contour coils, as well as anterior contour coils combined with spine coils, to compare the SNR in phantom and healthy subjects. Results: This research demonstrates that combining anterior and posterior contour coils doubles the SNR in abdominal MRI, enhancing image quality and supporting diagnostic accuracy. Impact: Using anterior and posterior contour coils improves imaging diagnostic accuracy in medical settings.
Authors
- Michael N. Hoff (ORCID: https://orcid.org/0000-0002-3498-0591)
- Emma Bahroos (ORCID: https://orcid.org/0000-0001-6542-9711)
- Michael A. Ohliger (ORCID: https://orcid.org/0000-0001-6878-8189)
- Vahid Ravanfar
- Yang Yang (ORCID: https://orcid.org/0000-0002-1732-3709)
- Daehyun Yoon
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/5393
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00