Assessing Blood-Brain Barrier Opening and Safety in Rats during Focused Ultrasound Procedure with MRI and Acoustic Cavitation Monitoring
Motivation: There is a lack of comprehensive MRI-based evaluation of low-intensity focused ultrasound (LIFU)-induced blood-brain barrier (BBB) opening in rat models. Goal(s): We aim to develop a safe BBB opening protocol in rat models with jugular vein catheter, using cavitation monitoring and MRI assessment. Approach: We investigated the effects of sonication parameters on BBB opening and tissue damage and the relationship between ultrasound cavitation measures and contrast changes on post-treatment MRI (T1W, T2W, SWI, T2*) Results: Optimal BBB opening was achieved at 0.45 MPa with 10 ms burst length and continuous microbubble injection, confirmed by T1-weighted MRI showing hyperintensity and no brain damage. Impact: The optimized BBB opening and monitoring protocol developed in this study will enable preclinical studies of delivery of the therapeutic compounds across the BBB.
Authors
- Kazim Narsinh (ORCID: https://orcid.org/0000-0002-2019-5461)
- Ki Soo Kim (ORCID: https://orcid.org/0000-0001-8387-9476)
- Eugene Ozhinsky (ORCID: https://orcid.org/0000-0001-8993-7905)
- Tianrun Xiao
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/2184
- Primary Topic
- Ultrasound and Hyperthermia Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00