Should I do 7T fMRI with the pTx or sTx coil? Consider the difference in geometry

Motivation: Parallel transmit (pTx) coils at 7T are clinically approved, but their advantage for fMRI remains unclear. Goal(s): We aimed to determine whether widely-used sTx or pTx coils provide better 7T fMRI data. Approach: We investigated 31 coils across 14 7T sites (SIEMENS and Philips), estimating noise correlation across receivers and mapped transmit field size/location using MR-visible beeswax and CT scans. Results: pTx coil placement benefits lower brain areas but increases noise coupling. Nova sTx and pTx coils differ in geometry. Impact: Geometrical design choices play a surprisingly large role for 7T fMRI when choosing the optimal coil.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/2333
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Should I do 7T fMRI with the pTx or sTx coil? Consider the difference in geometry

Anna I. Blazejewska, Wietske van der Zwaag, Luca Vizioli, Alessandra Pizzuti et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Should I do 7T fMRI with the pTx or sTx coil? Consider the difference in geometry

Anna I. Blazejewska, Wietske van der Zwaag, Luca Vizioli, Alessandra Pizzuti, Rüdiger Stirnberg, Essa Yacoub, Peter A. Bandettini, Nikos Priovoulos, Natalia Gudino, Dario Bosch, Jon̈athan R. Polimeni, Andrew Morgan, Monique C. Tourell, Joelle E. Sarlls, Barbara Dymerska, D. O. Gómez, Sebastian Dresbach, Laurentius Huber, Salvatore Torrisi, Yuhui Chai, An T. Vu, Hendrik Mattern, Benedikt Andreas Poser, Emma Brouwer, Jennifer Evans, Saskia Bollmann
article en

Abstract

Motivation: Parallel transmit (pTx) coils at 7T are clinically approved, but their advantage for fMRI remains unclear. Goal(s): We aimed to determine whether widely-used sTx or pTx coils provide better 7T fMRI data. Approach: We investigated 31 coils across 14 7T sites (SIEMENS and Philips), estimating noise correlation across receivers and mapped transmit field size/location using MR-visible beeswax and CT scans. Results: pTx coil placement benefits lower brain areas but increases noise coupling. Nova sTx and pTx coils differ in geometry. Impact: Geometrical design choices play a surprisingly large role for 7T fMRI when choosing the optimal coil.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 65%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.