Definition of the echo time for UTE imaging and short rectangular rf-pulses

Motivation: Currently, in UTE literature, there is inconsistency in defining echo time when short rectangular excitation pulses are used. Goal(s): Investigate whether echo-time should be defined from the center or the end of the short rectangular rf-pulse. Approach: Theoretical considerations, simulations, and measurements with a 3D-UTE sequence. Results: Two MR signal mechanisms were investigated: (I) relaxation and (II) off-resonance effects during the rf-pulse. In both cases, defining echo time from the rf-pulse center was more accurate. Impact: Results indicate that defining TE from the center of short rectangular RF pulses is more accurate, suggesting the community should adopt this TE definition for future UTE applications.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/2614
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Definition of the echo time for UTE imaging and short rectangular rf-pulses

Karl‐Heinz Herrmann, Jürgen Rainer Reichenbach, Martin Krämer
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Definition of the echo time for UTE imaging and short rectangular rf-pulses

Karl‐Heinz Herrmann, Jürgen Rainer Reichenbach, Martin Krämer
article en

Abstract

Motivation: Currently, in UTE literature, there is inconsistency in defining echo time when short rectangular excitation pulses are used. Goal(s): Investigate whether echo-time should be defined from the center or the end of the short rectangular rf-pulse. Approach: Theoretical considerations, simulations, and measurements with a 3D-UTE sequence. Results: Two MR signal mechanisms were investigated: (I) relaxation and (II) off-resonance effects during the rf-pulse. In both cases, defining echo time from the rf-pulse center was more accurate. Impact: Results indicate that defining TE from the center of short rectangular RF pulses is more accurate, suggesting the community should adopt this TE definition for future UTE applications.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 65%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Definition of the echo time for UTE imaging and short rectangular rf-pulses — Karl‐Heinz Herrmann, Jürgen Rainer Reichenbach, et al. · Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition (2025) | TGRS Research Map | TGRS