Fat-corrected APTw/CEST-MRI with timed breathing for renal applications
Motivation: APTw/CEST MRI in renal imaging would be useful for many ailments, such as characterization of small renal masses in cancer or kidney injury/transplants. However, fat and motion artifacts remain challenging. Goal(s): Assess feasibility of in vivo fat-correction in renal CEST-MRI using previously introduced Z-spectral fitting of SPIR fat-suppressed data. Approach: APTw/CEST data was acquired using single-shot fast-spin-echo MRI with compressed-sensing acceleration and timed-breathing in volunteers. Background model fitting (olefinic/aliphatic fat, water/MT, SPIR effect) is used for fat-correction in APTw imaging. Results: Initial renal CEST-MRI using Z-spectral background-fitting was demonstrated with good suppression of fat and detailed substructure of CEST effect in the organ. Impact: Enable CEST applications in the kidney with low artifact level from fat.
Authors
- Ivan Dimitrov (ORCID: https://orcid.org/0000-0003-2686-6884)
- Jochen Keupp (ORCID: https://orcid.org/0000-0003-2129-1153)
- Elena Vinogradov (ORCID: https://orcid.org/0000-0003-4607-4147)
- Holger Eggers
- Xiaojing Wang (ORCID: https://orcid.org/0000-0002-4049-8822)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2025-09-16
- DOI
- https://doi.org/10.58530/2025/2713
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00