Optimizing density of intrathecal contrast at digital subtraction myelography: Evaluation of differing needle characteristics and injection rates in a phantom
Dynamic myelography, performed as digital subtraction myelography or dynamic computed tomography myelography, is crucial in diagnosing intracranial hypotension resulting from a cerebrospinal fluid-venous fistula (CVF). The quality of the myelogram is paramount for accurate diagnosis. Using a phantom, the impact of needle type (Quincke vs. Whitacre), caliber, side-hole position, and rate of injection on the quality of the myelogram was determined. The ideal decubitus myelogram would provide a large volume of a hyperdense contrast within the lateral dependent aspect of the thecal sac, optimally flooding the mouths of the neural foramina and root sleeves where the vast majority of CVFs originate. The results of this study suggest it is exclusively the rate of injection that most predictably dictates the quality of the myelogram in this regard. Specifically, a slow injection rate, on the order of 0.1 mL/s, should be opted for to decrease turbulence, optimize myelogram quality, and thus improve CVF detection in clinical practice.
Authors
- Richard Farb (ORCID: https://orcid.org/0000-0002-7770-3762)
- Anahita Malvea (ORCID: https://orcid.org/0000-0001-6216-1510)
- Emily Chung (ORCID: https://orcid.org/0000-0002-1326-8693)
- Mehran Nasralla (ORCID: https://orcid.org/0000-0003-1754-2191)
- Eef Jacobus Hendriks (ORCID: https://orcid.org/0000-0003-2100-7992)
Institutions
- University of Toronto (CA)
Publication Details
- Journal
- Interventional Neuroradiology
- Published
- 2025-09-16
- DOI
- https://doi.org/10.1177/15910199251375541
- Primary Topic
- Neurosurgical Procedures and Complications
- Type
- article
- Field-Weighted Citation Impact
- 0.00