Active Eye Lens Dosimetry With Dosepix: Influence of Measurement Position and Lead Glass Shielding

In this work, the effect of the measurement position on measurements of Hp(3) of a new active eye lens dosemeter prototype based on the Dosepix detector is examined. A comparison between measuring directly in front of the eye and measuring at the side of the head of an Alderson phantom showed no significant influence on the resulting Hp(3) for different radiation qualities and angles. In addition, to account for the absorption effect of radiation safety glasses, pieces of lead glass were attached to the front and side of the dosemeter. The challenging geometry of the human face, when additionally equipped with absorbing glasses and irradiated at different angles of radiation incidence, led to a complex irradiation setup, whereby the dosemeter cannot be placed at the real point of interest, i.e. directly in front of the eye. Corresponding effects and consequences for radiation protection measurements have been investigated by using a human like Alderson head phantom as well as thermoluminescent dosemeters (TLDs) and the active eye lens dosemeter prototype based on the Dosepix detector. For specific angles, the radiation bypassed the radiation safety glasses and lead glass pieces, leading to an increase in the measured Hp(3). Compared to TLDs behind radiation safety glasses, measurements with the lead glass shielded prototype resulted in lower values for Hp(3). Furthermore, the results did not reproduce previous findings where larger dose values were found for Dosepix behind lead glass pieces than for the TLDs behind radiation safety glasses. A possible reason might be that the dimensions of the lead glass pieces are not representative of the radiation safety glasses in front of the eye but, ultimately, it is not yet clear what the main reason for the deviation is. Therefore, it is advisable to test the same methodology in future investigations with other eyewear models and lead glass pieces to investigate whether similar behaviors occur.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Radiation and Plasma Medical Sciences
Published
2025-09-16
DOI
https://doi.org/10.1109/trpms.2025.3609894
Primary Topic
Intraocular Surgery and Lenses
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Active Eye Lens Dosimetry With Dosepix: Influence of Measurement Position and Lead Glass Shielding

F. A. Beisser, Oliver Hupe, Hayo Zutz, R. Behrens et al.
IEEE Transactions on Radiation and Plasma Medical Sciences
Intraocular Surgery and Lenses
article

Active Eye Lens Dosimetry With Dosepix: Influence of Measurement Position and Lead Glass Shielding

F. A. Beisser, Oliver Hupe, Hayo Zutz, R. Behrens, Thilo Michel, G. Hilgers, S. Funk, Jürgen Roth, Leonie Ullmann, Tom Tröltzsch
article en

Abstract

In this work, the effect of the measurement position on measurements of Hp(3) of a new active eye lens dosemeter prototype based on the Dosepix detector is examined. A comparison between measuring directly in front of the eye and measuring at the side of the head of an Alderson phantom showed no significant influence on the resulting Hp(3) for different radiation qualities and angles. In addition, to account for the absorption effect of radiation safety glasses, pieces of lead glass were attached to the front and side of the dosemeter. The challenging geometry of the human face, when additionally equipped with absorbing glasses and irradiated at different angles of radiation incidence, led to a complex irradiation setup, whereby the dosemeter cannot be placed at the real point of interest, i.e. directly in front of the eye. Corresponding effects and consequences for radiation protection measurements have been investigated by using a human like Alderson head phantom as well as thermoluminescent dosemeters (TLDs) and the active eye lens dosemeter prototype based on the Dosepix detector. For specific angles, the radiation bypassed the radiation safety glasses and lead glass pieces, leading to an increase in the measured Hp(3). Compared to TLDs behind radiation safety glasses, measurements with the lead glass shielded prototype resulted in lower values for Hp(3). Furthermore, the results did not reproduce previous findings where larger dose values were found for Dosepix behind lead glass pieces than for the TLDs behind radiation safety glasses. A possible reason might be that the dimensions of the lead glass pieces are not representative of the radiation safety glasses in front of the eye but, ultimately, it is not yet clear what the main reason for the deviation is. Therefore, it is advisable to test the same methodology in future investigations with other eyewear models and lead glass pieces to investigate whether similar behaviors occur.

IEEE Transactions on Radiation and Plasma Medical Sciences
Physikalisch-Technische Bundesanstalt (DE), Friedrich-Alexander-Universität Erlangen-Nürnberg (DE)
Openalex Percentile: Top 63%
Intraocular Surgery and Lenses
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.