Accuracy And Reliability Of A Commercial Group Fitness Heart Rate Monitor Across Exercise Intensity

PURPOSE: To determine the accuracy and reliability of a commercial group fitness heart rate (HR) monitor during an incremental test to exhaustion on a cycle ergometer. METHODS: Twenty recreationally-trained, college-aged males (mean ± SD age = 21.1 ± 1.9 years; body mass = 84.7 ± 15.0 kg; height = 181.2 ± 7.5 cm; aerobic training = 1.8 ± 1.9 hr·wk-1; resistance training = 6.2 ± 2.7 hr·wk-1) volunteered to complete an incremental test to exhaustion on a cycle ergometer that began at 80 W and increased 30 W every two minutes. A group fitness heart rate monitor was placed on the forearm as well as the upper arm of each subject with a 12-lead electrocardiogram (ECG) used as the criterion method. Heart rate was recorded from each device every 30 seconds and averaged across each two-minute stage. Subjects were provided strong verbal encouragement throughout the test to achieve their maximal exercise intensity (i.e. within 10 bpm of age-predicted maximum HR and rating of perceived exertion ≥18). Accuracy of each heart rate monitor was assessed through calculation of mean absolute percent error (MAPE), simple linear regression, and Bland-Altman plot. An intraclass correlation coefficient (ICC) was used to assess inter-device reliability. Statistical significance was set at an alpha level of 0.05. RESULTS: The MAPE (± SD) values were 0.9 (± 0.6)% and 0.8 (± 0.5)% for the forearm and the upper arm heart rate monitors, respectively. Linear regression analyses for ECG HR vs. forearm HR resulted in a strong correlation (r = 0.998) and standard error of estimate (SEE) value of 1.2 bpm, whereas ECG HR vs. upper arm HR resulted in a strong correlation (r = 0.999) and SEE value of 1.0 bpm. The mean (± SD) constant error (CE) (actual ECG HR - predicted HR) values were 1.0 (± 1.2) bpm and 0.8 (± 1.1) bpm for the forearm and upper arm, respectively. The Bland-Altman regression analyses (CE vs. ECG HR) resulted in a non-significant correlation for the forearm (r = 0.107, p = 0.268), but a significant correlation for the upper arm (r = 0.303, p < 0.001). The ICC for inter-device reliability between the forearm and upper arm monitors was R = 0.999. CONCLUSIONS: Our findings indicated the group fitness heart rate monitors placed on the forearm and upper arm provided highly accurate and reliable values when compared to an ECG from low to maximal exercise intensities.

Authors

Institutions

Publication Details

Journal
Medicine & Science in Sports & Exercise
Published
2025-09-16
DOI
https://doi.org/10.1249/01.mss.0001160208.72055.23
Primary Topic
Cardiovascular and exercise physiology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy And Reliability Of A Commercial Group Fitness Heart Rate Monitor Across Exercise Intensity

Christopher M. Hill, Matthew P. Herring, Clayton L. Camic, Matteo F. de Leon et al.
Medicine & Science in Sports & Exercise
Cardiovascular and exercise physiology
article

Accuracy And Reliability Of A Commercial Group Fitness Heart Rate Monitor Across Exercise Intensity

Christopher M. Hill, Matthew P. Herring, Clayton L. Camic, Matteo F. de Leon, Ethan G. Wunner
article en

Abstract

PURPOSE: To determine the accuracy and reliability of a commercial group fitness heart rate (HR) monitor during an incremental test to exhaustion on a cycle ergometer. METHODS: Twenty recreationally-trained, college-aged males (mean ± SD age = 21.1 ± 1.9 years; body mass = 84.7 ± 15.0 kg; height = 181.2 ± 7.5 cm; aerobic training = 1.8 ± 1.9 hr·wk-1; resistance training = 6.2 ± 2.7 hr·wk-1) volunteered to complete an incremental test to exhaustion on a cycle ergometer that began at 80 W and increased 30 W every two minutes. A group fitness heart rate monitor was placed on the forearm as well as the upper arm of each subject with a 12-lead electrocardiogram (ECG) used as the criterion method. Heart rate was recorded from each device every 30 seconds and averaged across each two-minute stage. Subjects were provided strong verbal encouragement throughout the test to achieve their maximal exercise intensity (i.e. within 10 bpm of age-predicted maximum HR and rating of perceived exertion ≥18). Accuracy of each heart rate monitor was assessed through calculation of mean absolute percent error (MAPE), simple linear regression, and Bland-Altman plot. An intraclass correlation coefficient (ICC) was used to assess inter-device reliability. Statistical significance was set at an alpha level of 0.05. RESULTS: The MAPE (± SD) values were 0.9 (± 0.6)% and 0.8 (± 0.5)% for the forearm and the upper arm heart rate monitors, respectively. Linear regression analyses for ECG HR vs. forearm HR resulted in a strong correlation (r = 0.998) and standard error of estimate (SEE) value of 1.2 bpm, whereas ECG HR vs. upper arm HR resulted in a strong correlation (r = 0.999) and SEE value of 1.0 bpm. The mean (± SD) constant error (CE) (actual ECG HR - predicted HR) values were 1.0 (± 1.2) bpm and 0.8 (± 1.1) bpm for the forearm and upper arm, respectively. The Bland-Altman regression analyses (CE vs. ECG HR) resulted in a non-significant correlation for the forearm (r = 0.107, p = 0.268), but a significant correlation for the upper arm (r = 0.303, p < 0.001). The ICC for inter-device reliability between the forearm and upper arm monitors was R = 0.999. CONCLUSIONS: Our findings indicated the group fitness heart rate monitors placed on the forearm and upper arm provided highly accurate and reliable values when compared to an ECG from low to maximal exercise intensities.

Medicine & Science in Sports & ExerciseVol. 57(10S)
Northern Illinois University (US), Louisiana State University (US)
Openalex Percentile: Top 61%
Cardiovascular and exercise physiology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.