Optimization and comparison of [18F]FET synthesis on two distinct automated radiochemistry systems
<i>O</i>-(2-[<sup>18</sup>F]Fluoroethyl)- <i><sub>L</sub></i> -tyrosine ([<sup>18</sup>F]FET) is a promising amino acid PET tracer for assessment of malignant brain tumors. Herein, we report optimized production of [<sup>18</sup>F]FET for clinical use on two commercial radiochemistry systems: Sofie ELIXYS and GE FASTlab 2. While the Sofie ELIXYS procedure requires high performance liquid chromatography (HPLC) purification, the GE FASTlab 2 method uses solid-phase extraction (SPE) purification. In both cases, [<sup>18</sup>F]FET met release specifications for clinical investigation laid out in the United States Pharmacopeia (USP) and/or European Pharmacopeia (Ph. Eur.). The radiochemical yield of [<sup>18</sup>F]FET was 35-55% and 30-55% decay corrected to start of synthesis (SOS) for Sofie ELIXYS and GE FASTlab 2, respectively. The overall synthesis time was 75-85 and 70-80 min from SOS for Sofie ELIXYS and GE FASTlab 2, respectively. The radiochemical purity was > 99%, and the molar activity (A<sub>m</sub>) was 340-464 GBq/µmol at end of synthesis (EOS).
Authors
- Qi‐Huang Zheng (ORCID: https://orcid.org/0000-0001-6455-8423)
Institutions
- Indiana University – Purdue University Indianapolis (US)
Publication Details
- Journal
- American Journal of Nuclear Medicine and Molecular Imaging
- Published
- 2025-01-01
- DOI
- https://doi.org/10.62347/spoe8395
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- National Institutes of Health
- National Center for Advancing Translational Sciences