Phase‐Contrast MRI With Hybrid One‐ and Two‐Sided Simultaneous Three‐Directional Flow‐Encoding and Velocity Spectrum Separation (HOTSPA+)

A phase-contrast MRI (PC-MRI) acceleration technique was developed using hybrid one- and two-sided simultaneous three-directional flow-encoding and velocity spectrum separation (HOTSPA+). HOTSPA+ simultaneously encodes blood flow in three directions, using one-sided encoding for the z direction and two-sided encoding for both the x- and y-directions. By this encoding method, spectra of blood flow velocities in three directions can be separated from each other by temporal frequency filtering, accelerating the acquisition, and improving the flow temporal resolution. To demonstrate the benefits of HOTSPA+, common carotid data were acquired from 10 healthy volunteers and compared with the standard PC-MRI and HOTSPA, which is a previously proposed technique that can simultaneously encode flow up to two directions. Compared to the standard PC-MRI, acquisition time using HOTSPA+ was reduced by threefold, while the flow measurements showed strong correlations with R<sup>2</sup> = 0.969 for the total volumetric flow and R<sup>2</sup> = 0.936 for the peak velocity. The total volumetric flow showed a bias of -0.09 mL (relative bias error of 3.18%), and the peak velocity demonstrated a bias of -0.16 cm/s (relative bias error of 0.18%). When using the same acquisition time, HOTSPA+ acquired higher peak velocity and more accurate mean velocity than the standard PC-MRI, while measuring similar total volumetric flow (R<sup>2</sup> = 0.971, bias = -0.06 mL, relative bias error = 1.19%), indicating that HOTSPA+ can also have higher flow temporal resolution than conventional methods. By simultaneously encoding the flow in three directions and using frequency filtering to separate and recover the velocities, our proposed HOTSPA+ technique can significantly reduce the scan time while accurately measuring the peak flow velocity and total volumetric flow.

Authors

Institutions

Publication Details

Journal
NMR in Biomedicine
Published
2025-09-12
DOI
https://doi.org/10.1002/nbm.70145
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Phase‐Contrast MRI With Hybrid One‐ and Two‐Sided Simultaneous Three‐Directional Flow‐Encoding and Velocity Spectrum Separation (HOTSPA+)

Zijian Zhou, Wenjian Liu, Haikun Qi, Junpu Hu et al.
NMR in Biomedicine
Advanced MRI Techniques and Applications
article

Phase‐Contrast MRI With Hybrid One‐ and Two‐Sided Simultaneous Three‐Directional Flow‐Encoding and Velocity Spectrum Separation (HOTSPA+)

Zijian Zhou, Wenjian Liu, Haikun Qi, Junpu Hu, Jian Xu, Peng Hu, Jiayu Zhu
article en

Abstract

A phase-contrast MRI (PC-MRI) acceleration technique was developed using hybrid one- and two-sided simultaneous three-directional flow-encoding and velocity spectrum separation (HOTSPA+). HOTSPA+ simultaneously encodes blood flow in three directions, using one-sided encoding for the z direction and two-sided encoding for both the x- and y-directions. By this encoding method, spectra of blood flow velocities in three directions can be separated from each other by temporal frequency filtering, accelerating the acquisition, and improving the flow temporal resolution. To demonstrate the benefits of HOTSPA+, common carotid data were acquired from 10 healthy volunteers and compared with the standard PC-MRI and HOTSPA, which is a previously proposed technique that can simultaneously encode flow up to two directions. Compared to the standard PC-MRI, acquisition time using HOTSPA+ was reduced by threefold, while the flow measurements showed strong correlations with R<sup>2</sup> = 0.969 for the total volumetric flow and R<sup>2</sup> = 0.936 for the peak velocity. The total volumetric flow showed a bias of -0.09 mL (relative bias error of 3.18%), and the peak velocity demonstrated a bias of -0.16 cm/s (relative bias error of 0.18%). When using the same acquisition time, HOTSPA+ acquired higher peak velocity and more accurate mean velocity than the standard PC-MRI, while measuring similar total volumetric flow (R<sup>2</sup> = 0.971, bias = -0.06 mL, relative bias error = 1.19%), indicating that HOTSPA+ can also have higher flow temporal resolution than conventional methods. By simultaneously encoding the flow in three directions and using frequency filtering to separate and recover the velocities, our proposed HOTSPA+ technique can significantly reduce the scan time while accurately measuring the peak flow velocity and total volumetric flow.

NMR in BiomedicineVol. 38(10)
University of America (US), ShanghaiTech University (CN), United Imaging Healthcare (China) (CN)
Openalex Percentile: Top 66%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.