Quality control of plastic scintillators for the total-body J-PET scanner

Abstract The construction of the total-body Jagiellonian positron emission tomography scanner requires component verification before detector assembly. The purpose of this research is to verify the quality of BC-408 plastic scintillators with dimensions 6 mm × 30 mm × 330 mm. The scintillators were inspected for optical and mechanical defects and all dimensions were measured. Scratches, mechanically damaged corners and edges, as well as encapsulated dust and fibers within the scintillators, were identified under ceiling lamp illumination. Line defects on the as-cast surfaces were easily visible in a plane polariscope setup consisting of crossed horizontal and vertical polarizer foils. The transmittance at the wavelength of maximum emission through 6 mm thick scintillator samples and the technical attenuation length along 330 mm long scintillator samples were measured on a linear CCD array spectrometer for randomly selected scintillators from each delivered batch. Selected properties of the emission spectra, such as their FWHM and the values of the emission maxima as a function of the distance between the excitation point and the spectrometer fiber, were measured. Additionally, the scintillators optical homogeneity was measured on a light transfer setup consisting of an LED and photodiode matrix. The majority of the obtained plastic scintillators meets the transparency criteria and falls within the dimensional tolerances.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2025-09-01
DOI
https://doi.org/10.1088/1748-0221/20/09/p09019
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quality control of plastic scintillators for the total-body J-PET scanner

T. Kaplanoglu, D. Kumar, Catalina Curceanu, M. Skurzok et al.
1 citations
Journal of Instrumentation
Medical Imaging Techniques and Applications
1.36
article

Quality control of plastic scintillators for the total-body J-PET scanner

T. Kaplanoglu, D. Kumar, Catalina Curceanu, M. Skurzok, E. Czerwiński, Edward Lisowski, Piyush Pandey, Szymon Niedźwiecki, Paweł Moskal, E. Y. Beyene, K. Eliyan, Karol Kubat, Manish Das, Gabriel Moskal, S. Parzych, P. Tanty, Ł. Kapłon, K. Tayefi, N. Chug, E. Pérez del Río, K. Kacprzak, G. Korcyl, Ewa Łucja Stępień, W. Mryka, S. Sharma, Filip Lisowski, J. Mędrala-Sowa, Anoop Kunimmal Venadan
article en
1 citations

Abstract

Abstract The construction of the total-body Jagiellonian positron emission tomography scanner requires component verification before detector assembly. The purpose of this research is to verify the quality of BC-408 plastic scintillators with dimensions 6 mm × 30 mm × 330 mm. The scintillators were inspected for optical and mechanical defects and all dimensions were measured. Scratches, mechanically damaged corners and edges, as well as encapsulated dust and fibers within the scintillators, were identified under ceiling lamp illumination. Line defects on the as-cast surfaces were easily visible in a plane polariscope setup consisting of crossed horizontal and vertical polarizer foils. The transmittance at the wavelength of maximum emission through 6 mm thick scintillator samples and the technical attenuation length along 330 mm long scintillator samples were measured on a linear CCD array spectrometer for randomly selected scintillators from each delivered batch. Selected properties of the emission spectra, such as their FWHM and the values of the emission maxima as a function of the distance between the excitation point and the spectrometer fiber, were measured. Additionally, the scintillators optical homogeneity was measured on a light transfer setup consisting of an LED and photodiode matrix. The majority of the obtained plastic scintillators meets the transparency criteria and falls within the dimensional tolerances.

Journal of InstrumentationVol. 20(09)
Jagiellonian University (PL), Cracow University of Technology (PL), Istituto Nazionale di Fisica Nucleare, Laboratori Nazionali di Frascati (IT)
Openalex Percentile: Top 16%
Medical Imaging Techniques and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.