The effect of different height titanium base abutments on the internal fit of a zirconia crown

Background and objectives. This in vitro study aimed to evaluate the effect of titanium base abutment height on the internal fit of CAD/CAM-fabricated zirconia crowns. Achieving a precise internal fit is critical to the long-term success of implant-supported restorations, yet limited data exist regarding how abutment height may influence this parameter. Materials and methods. Twenty implant analogues made of Grade 4 titanium (diameters: 4.5 mm and 3.75 mm; lengths: 12 mm and 13 mm) and twenty titanium base abutments (diameter: 3 mm) with two different sulcular heights (6 mm and 4 mm), along with their corresponding scan bodies, were selected. The specimens were divided into two groups of 10 each. Group A employed Dentium® scan bodies and analogues with 6 mm abutment height, while Group B used SGS® scan bodies and analogues with 4 mm abutment height. Crown design and milling were performed using a standardized CAD/CAM workflow. Internal fit was assessed using the digital silicone replica technique and 3D analysis with Geomagic Control X software, both before and after thermocycling. Mechanical performance was evaluated by measuring removal torque values (RTVs) before and after cyclic loading. Results. There were no statistically significant differences in internal fit between the two groups, both before (p = 0.605) and after thermocycling (p = 0.800). Similarly, no significant differences were found in pre- and post-loading RTVs between groups (p > 0.05). Conclusion. The findings of this research suggest, with certain caveats, that titanium base abutments may serve as a viable alternative to standard dental implant restoration procedures. Careful selection and customization of the abutment height should be tailored to the specific clinical case, taking into account various factors such as crown dimensions, patient-specific anatomical considerations, and the functional demands of the restoration.

Authors

Publication Details

Journal
Romanian Journal of Stomatology
Published
2025-09-10
DOI
https://doi.org/10.37897/rjs.2025.3.3
Primary Topic
Bone Tissue Engineering Materials
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The effect of different height titanium base abutments on the internal fit of a zirconia crown

Firas Saddam Oglah Albaaj, Ahmed A. Jasim, Nada Ali Mahdi, Samah F. Mohamed‐Ali
Romanian Journal of Stomatology
Bone Tissue Engineering Materials
article

The effect of different height titanium base abutments on the internal fit of a zirconia crown

Firas Saddam Oglah Albaaj, Ahmed A. Jasim, Nada Ali Mahdi, Samah F. Mohamed‐Ali
article en

Abstract

Background and objectives. This in vitro study aimed to evaluate the effect of titanium base abutment height on the internal fit of CAD/CAM-fabricated zirconia crowns. Achieving a precise internal fit is critical to the long-term success of implant-supported restorations, yet limited data exist regarding how abutment height may influence this parameter. Materials and methods. Twenty implant analogues made of Grade 4 titanium (diameters: 4.5 mm and 3.75 mm; lengths: 12 mm and 13 mm) and twenty titanium base abutments (diameter: 3 mm) with two different sulcular heights (6 mm and 4 mm), along with their corresponding scan bodies, were selected. The specimens were divided into two groups of 10 each. Group A employed Dentium® scan bodies and analogues with 6 mm abutment height, while Group B used SGS® scan bodies and analogues with 4 mm abutment height. Crown design and milling were performed using a standardized CAD/CAM workflow. Internal fit was assessed using the digital silicone replica technique and 3D analysis with Geomagic Control X software, both before and after thermocycling. Mechanical performance was evaluated by measuring removal torque values (RTVs) before and after cyclic loading. Results. There were no statistically significant differences in internal fit between the two groups, both before (p = 0.605) and after thermocycling (p = 0.800). Similarly, no significant differences were found in pre- and post-loading RTVs between groups (p > 0.05). Conclusion. The findings of this research suggest, with certain caveats, that titanium base abutments may serve as a viable alternative to standard dental implant restoration procedures. Careful selection and customization of the abutment height should be tailored to the specific clinical case, taking into account various factors such as crown dimensions, patient-specific anatomical considerations, and the functional demands of the restoration.

Romanian Journal of StomatologyVol. 71(3)
Openalex Percentile: Top 77%
Bone Tissue Engineering Materials
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

The effect of different height titanium base abutments on the internal fit of a zirconia crown — Firas Saddam Oglah Albaaj, Ahmed A. Jasim, et al. · Romanian Journal of Stomatology (2025) | TGRS Research Map | TGRS