Clinical comparison of four commercial MRI phantoms on a 0.35T MR‐linac using a standard clinical TRUFI sequence at a single institution
All four phantoms successfully measured distortion at multiple gantry angles on a 0.35 T MR-Linac. The QUASAR MRID<sup>3D</sup> Phantom uniquely differentiated B<sub>0</sub> from gradient distortions but required an offset adjustment. The MagPhan and QUASAR Insight phantoms offered comprehensive imaging metrics, while the Sun Nuclear Phantom exhibited relatively higher distortions.