A comprehensive investigation of the radiation isocentre spatial variability in linear accelerators: implications for commissioning, QA, and clinical protocols

Stereotactic ablative radiation therapy and stereotactic radiosurgery (SRS) are modern high precision techniques used for cancer management. Hence, evaluating potential clinical impacts of mischaracterisation of the radiation isocentre (RI) variability at different gantry and collimator angles during commissioning and quality assurance (QA) helps to ensure the robustness of treatment delivery and successful outcomes. This was done using signal edge detection techniques on flattened and flattening filter free (FFF) fields generated by megavoltage (MV) treatment beams. A total of 22,815 MV images of 4 × 16 cm2, were captured using an iViewGT™ Perkin Elmer a-Si Electronic Portal Imaging Device at 13 different gantry angles, paired with 13 collimator angles each in 30° intervals. 6 MV, 6 MV FFF, and 18 MV photon treatment modes were used across 8 linacs. The Starshot images were subsequently analysed using custom in-house signal detection software developed in Python. The mean RI variabilities across the entire treatment arc were approximately 0.58 mm, 0.66 mm, and 0.58 mm, for beam energies of 6 MV, 6 MV FFF, and 18 MV respectively, with corresponding standard deviations of 0.18 mm, 0.21 mm, and 0.19 mm, with maximum variabilities occurring at 300°, 0°, and 240° of magnitudes 0.62 mm, 0.71 mm, and 0.62 mm, respectively. Correspondingly, the standard deviations were 0.20 mm, 0.23 mm, and 0.17 mm. The resulting RI variabilities, applied to SRS QA parameters, may breach the 1 mm tolerance. In some clinical scenarios, this may impact treatment plan acceptance with treatment coverage goals outside of acceptable limits. We recommend enhancing commissioning and QA protocols by combining Starshot and Winston–Lutz-style tests to assess the positional deviation of the apparent radiation isocentre from the mechanical isocentre, and the radiation isocentre’s maximal variability across a larger range of intermediate angles. This approach helps define reference angles and tailor QA frequency based on individual linac behaviour.

Authors

Institutions

Publication Details

Journal
Physical and Engineering Sciences in Medicine
Published
2025-09-10
DOI
https://doi.org/10.1007/s13246-025-01637-8
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A comprehensive investigation of the radiation isocentre spatial variability in linear accelerators: implications for commissioning, QA, and clinical protocols

Pejman Rowshanfarzad, H.L. Riis, Thomas Milan, Z. Chen et al.
Physical and Engineering Sciences in Medicine
Advanced Radiotherapy Techniques
article

A comprehensive investigation of the radiation isocentre spatial variability in linear accelerators: implications for commissioning, QA, and clinical protocols

Pejman Rowshanfarzad, H.L. Riis, Thomas Milan, Z. Chen, Rohen White
article en

Abstract

Stereotactic ablative radiation therapy and stereotactic radiosurgery (SRS) are modern high precision techniques used for cancer management. Hence, evaluating potential clinical impacts of mischaracterisation of the radiation isocentre (RI) variability at different gantry and collimator angles during commissioning and quality assurance (QA) helps to ensure the robustness of treatment delivery and successful outcomes. This was done using signal edge detection techniques on flattened and flattening filter free (FFF) fields generated by megavoltage (MV) treatment beams. A total of 22,815 MV images of 4 × 16 cm2, were captured using an iViewGT™ Perkin Elmer a-Si Electronic Portal Imaging Device at 13 different gantry angles, paired with 13 collimator angles each in 30° intervals. 6 MV, 6 MV FFF, and 18 MV photon treatment modes were used across 8 linacs. The Starshot images were subsequently analysed using custom in-house signal detection software developed in Python. The mean RI variabilities across the entire treatment arc were approximately 0.58 mm, 0.66 mm, and 0.58 mm, for beam energies of 6 MV, 6 MV FFF, and 18 MV respectively, with corresponding standard deviations of 0.18 mm, 0.21 mm, and 0.19 mm, with maximum variabilities occurring at 300°, 0°, and 240° of magnitudes 0.62 mm, 0.71 mm, and 0.62 mm, respectively. Correspondingly, the standard deviations were 0.20 mm, 0.23 mm, and 0.17 mm. The resulting RI variabilities, applied to SRS QA parameters, may breach the 1 mm tolerance. In some clinical scenarios, this may impact treatment plan acceptance with treatment coverage goals outside of acceptable limits. We recommend enhancing commissioning and QA protocols by combining Starshot and Winston–Lutz-style tests to assess the positional deviation of the apparent radiation isocentre from the mechanical isocentre, and the radiation isocentre’s maximal variability across a larger range of intermediate angles. This approach helps define reference angles and tailor QA frequency based on individual linac behaviour.

Physical and Engineering Sciences in MedicineVol. 48(4)
The University of Western Australia (AU), University of Southern Denmark (DK), Government of Western Australia Department of Health (AU), Odense University Hospital (DK), Sir Charles Gairdner Hospital (AU), Westmead Hospital (AU)
Odense Universitetshospital
Openalex Percentile: Top 69%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.