3D isotropic FastView MRI localizer allows reliable torsion measurements of the lower limb

Computed tomography (CT) and magnetic resonance imaging (MRI) are commonly used to assess femoral and tibial torsion. While CT offers high spatial resolution, it involves ionizing radiation. MRI avoids radiation but requires multiple sequences and extended acquisition time. We retrospectively evaluated whether a three-dimensional isotropic MRI localizer (FastView) could serve as a reliable and faster alternative. In this retrospective single-center study, 60 lower limbs from 30 patients, aged 27.1 ± 11.5 years (mean ± standard deviation), 19 females and 11 males, were assessed using both FastView and a dedicated MRI protocol. FastView (5 × 5 × 5 mm<sup>3</sup> voxels) imaged the entire lower limb in 17.4 s compared to nearly 7 min for the dedicated protocol. Torsion angles were measured independently by two readers. Agreement between methods was evaluated using intraclass correlation coefficients (ICCs), Bland-Altman plots, and Pearson R². No significant differences in torsion values were found (all p > 0.305). Femoral (ICC: 0.91-0.96) and tibial (ICC: 0.91-0.94) torsion showed excellent inter-modality agreement. Inter-reader reliability was also high (ICC: 0.95-0.99). Correlation values confirmed strong agreement (R²: 0.891-0.963). FastView demonstrated accuracy comparable to the dedicated protocol, offering a fast, efficient, and radiation-free option for routine torsion assessment. RELEVANCE STATEMENT: FastView MRI localizer offers a fast and resource-efficient method for assessing lower limb torsion, potentially replacing standard multisequence protocols in routine clinical practice. KEY POINTS: FastView MRI enables lower limb torsion measurements with full-limb coverage in under 20 s. Torsion angles from FastView and dedicated MRI showed no significant differences. Femoral and tibial ICCs between 0.91 and 0.96 confirm excellent inter-protocol agreement. Inter-reader agreement was consistently high across both protocols. FastView may replace multisequence MRI protocols in routine clinical torsion assessment.

Authors

Institutions

Publication Details

Journal
European Radiology Experimental
Published
2025-09-09
DOI
https://doi.org/10.1186/s41747-025-00631-9
Citations
1
Primary Topic
Hip disorders and treatments
Type
article
Field-Weighted Citation Impact
2.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

3D isotropic FastView MRI localizer allows reliable torsion measurements of the lower limb

Nina Hesse, Lennart Schröder, Boj Hoppe, Hannah Gildein et al.
1 citations
European Radiology Experimental
Hip disorders and treatments
2.01
article

3D isotropic FastView MRI localizer allows reliable torsion measurements of the lower limb

Nina Hesse, Lennart Schröder, Boj Hoppe, Hannah Gildein, Jens Ricke, Volker Schäfer, Boris Michael Holzapfel, Paul Reidler, Friederike Herr, Natascha Hohmann, Cristina Mihaela Dascalescu, Jörg Arnholdt
article en
1 citations

Abstract

Computed tomography (CT) and magnetic resonance imaging (MRI) are commonly used to assess femoral and tibial torsion. While CT offers high spatial resolution, it involves ionizing radiation. MRI avoids radiation but requires multiple sequences and extended acquisition time. We retrospectively evaluated whether a three-dimensional isotropic MRI localizer (FastView) could serve as a reliable and faster alternative. In this retrospective single-center study, 60 lower limbs from 30 patients, aged 27.1 ± 11.5 years (mean ± standard deviation), 19 females and 11 males, were assessed using both FastView and a dedicated MRI protocol. FastView (5 × 5 × 5 mm<sup>3</sup> voxels) imaged the entire lower limb in 17.4 s compared to nearly 7 min for the dedicated protocol. Torsion angles were measured independently by two readers. Agreement between methods was evaluated using intraclass correlation coefficients (ICCs), Bland-Altman plots, and Pearson R². No significant differences in torsion values were found (all p > 0.305). Femoral (ICC: 0.91-0.96) and tibial (ICC: 0.91-0.94) torsion showed excellent inter-modality agreement. Inter-reader reliability was also high (ICC: 0.95-0.99). Correlation values confirmed strong agreement (R²: 0.891-0.963). FastView demonstrated accuracy comparable to the dedicated protocol, offering a fast, efficient, and radiation-free option for routine torsion assessment. RELEVANCE STATEMENT: FastView MRI localizer offers a fast and resource-efficient method for assessing lower limb torsion, potentially replacing standard multisequence protocols in routine clinical practice. KEY POINTS: FastView MRI enables lower limb torsion measurements with full-limb coverage in under 20 s. Torsion angles from FastView and dedicated MRI showed no significant differences. Femoral and tibial ICCs between 0.91 and 0.96 confirm excellent inter-protocol agreement. Inter-reader agreement was consistently high across both protocols. FastView may replace multisequence MRI protocols in routine clinical torsion assessment.

European Radiology ExperimentalVol. 9(1)
LMU Klinikum (DE), Ludwig-Maximilians-Universität München (DE)
Openalex Percentile: Top 12%
Hip disorders and treatments
2.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.