Comparison of 30-2 Visual Field Using Melbourne Rapid Fields Online Perimetry and Humphrey Field Analyzer
MRF provides a portable and accessible alternative to HFA for 30-2 visual field testing, with good agreement in moderate to advanced glaucoma. However, its slightly higher false negatives, longer test duration, and systemic difference in output to HFA should be considered when interpreting results. Further improvements may enhance its clinical utility.