Structural integrity and failure of transfemoral prosthetic socket fabricated using carbon prepreg technique: Influence of fiber orientation and curing conditions

This study provides valuable guidance for simplifying fabrication procedures and enhancing the structural integrity and safety of carbon fiber (CF) laminate transfemoral (TF) prosthetic sockets. While the high specific strength of CF laminate sockets offers advantages over conventional plastics, essential production data-their orientation-dependent strength and optimal cure conditions-are lacking, often requiring complex, costly cure cycles. This study investigated (i) the influence of fiber orientation on TF prosthetic CF socket strength via finite element analysis (FEA) during standing, and (ii) optimal single-step Vacuum-Bag-Only (VBO) cure conditions for prepreg in a low-cost conventional oven. Three distinct CF laminates ((45/-45/45/-45), (0/90/0/90), (0/45/-45/90)) were implemented in TF socket finite element (FE) models. Tensile and flexural tests validated FE results and assessed laminate failure modes. Differential Scanning Calorimetry (DSC) investigated cure temperatures, while surface voids were inspected to identify optimal single-step cure conditions. A 1-h isothermal cure at 90°C facilitated resin flow and yielded minimal surface voids. FEA revealed ply orientations insignificantly influenced residual limb pressure. Most plies in the (45/-45/45/-45) CF laminate favorably aligned with oblique deformation for TF socket stabilization during standing. Experimentally, it exhibited the lowest stiffness (10.86 GPa) and strength (161.49 MPa). Nevertheless, its strength is superior to other socket materials and enhances safety through clear pre-fracture signs from ductile failure. Maximum pressure of up to 32.2 kPa at the medial-distal site during standing was insufficient to cause discomfort. These findings provide guidelines for high-quality TF sockets using prepreg by simplifying the fabrication process.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2025-09-01
DOI
https://doi.org/10.1177/09544119251372313
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Structural integrity and failure of transfemoral prosthetic socket fabricated using carbon prepreg technique: Influence of fiber orientation and curing conditions

Kazuhiko Sasaki, Pimpet Sratong-on, Watcharin Noothong, Chuen Kum Lee et al.
5 links
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Orthopaedic implants and arthroplasty
article

Structural integrity and failure of transfemoral prosthetic socket fabricated using carbon prepreg technique: Influence of fiber orientation and curing conditions

Kazuhiko Sasaki, Pimpet Sratong-on, Watcharin Noothong, Chuen Kum Lee, Jongsak Pakpia, Muhammad Aniq Bin Rezaki, Natasha Shong Wey Mun, Sawanya Suwannawong
article en

Abstract

This study provides valuable guidance for simplifying fabrication procedures and enhancing the structural integrity and safety of carbon fiber (CF) laminate transfemoral (TF) prosthetic sockets. While the high specific strength of CF laminate sockets offers advantages over conventional plastics, essential production data-their orientation-dependent strength and optimal cure conditions-are lacking, often requiring complex, costly cure cycles. This study investigated (i) the influence of fiber orientation on TF prosthetic CF socket strength via finite element analysis (FEA) during standing, and (ii) optimal single-step Vacuum-Bag-Only (VBO) cure conditions for prepreg in a low-cost conventional oven. Three distinct CF laminates ((45/-45/45/-45), (0/90/0/90), (0/45/-45/90)) were implemented in TF socket finite element (FE) models. Tensile and flexural tests validated FE results and assessed laminate failure modes. Differential Scanning Calorimetry (DSC) investigated cure temperatures, while surface voids were inspected to identify optimal single-step cure conditions. A 1-h isothermal cure at 90°C facilitated resin flow and yielded minimal surface voids. FEA revealed ply orientations insignificantly influenced residual limb pressure. Most plies in the (45/-45/45/-45) CF laminate favorably aligned with oblique deformation for TF socket stabilization during standing. Experimentally, it exhibited the lowest stiffness (10.86 GPa) and strength (161.49 MPa). Nevertheless, its strength is superior to other socket materials and enhances safety through clear pre-fracture signs from ductile failure. Maximum pressure of up to 32.2 kPa at the medial-distal site during standing was insufficient to cause discomfort. These findings provide guidelines for high-quality TF sockets using prepreg by simplifying the fabrication process.

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in MedicineVol. 239(9)
Siriraj Hospital (TH), Singapore Institute of Technology (SG), Nanyang Polytechnic (SG), Thai-Nichi Institute of Technology (TH)
Takahashi Industrial and Economic Research Foundation, Thai-Nichi Institute of Technology
Openalex Percentile: Top 62%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.