ToF-SIMS spectra of typical substrates in both polarities: Si wafer, Au-coated glass, and ITO-coated glass
ToF-SIMS is a surface analysis technique, and as a result, many samples need to be spotted or mounted onto a suitable substrate. Different substrates can be used for ToF-SIMS analysis, and it is essential to properly assign the characteristic chemical signals from the substrate to distinguish them from the sample analyte ions. Here, the authors provide a reference database that provides ToF-SIMS spectra for three widely used substrates in positive and negative polarities. The six spectra were acquired using a 30 keV Bi3+ primary ion beam, with Ar gas cluster ion beam sputter cleaning of the surfaces before analysis.
Authors
- Sebastiaan Van Nuffel (ORCID: https://orcid.org/0000-0003-1251-3236)
- Kimberly G. Garcia (ORCID: https://orcid.org/0000-0001-6187-7744)
- Eloise Gaspar
Institutions
- Maastricht University (NL)
- Department of Health (TW)
Publication Details
- Journal
- Surface Science Spectra
- Published
- 2025-09-05
- DOI
- https://doi.org/10.1116/6.0004796
- Primary Topic
- Ion-surface interactions and analysis
- Type
- article
- Field-Weighted Citation Impact
- 0.00