Simultaneous Detection and Differentiation of SARS-CoV-2, Influenza A/B, and Respiratory Syncytial Viruses in Respiratory Specimens Using the VitaSIRO solo™ SARS-CoV-2/Flu/RSV Assay

<b>Background/Objectives:</b> The concurrent circulation of SARS-CoV-2 with influenza A and B viruses and respiratory syncytial virus (RSV) represents a new diagnostic challenge in the post-COVID-19 area, especially considering that these infections have overlapping clinical presentations but different approaches to treatment and management. Multiplexed molecular testing on point-of-care platforms that focus on the simultaneous detection of multiple respiratory viruses in a single tube constitutes a useful approach for diagnosis of respiratory infections in decentralized clinical settings. This study evaluated the analytical performances of the VitaSIRO <i>solo</i>™ SARS-CoV-2/Flu/RSV Assay performed on the VitaSIRO <i>solo</i>™ Instrument (Credo Diagnostics Biomedical Pte. Ltd., Singapore, Republic of Singapore). <b>Methods:</b> With a view to accreditation, the criteria of the 2022-revised EN ISO 15189:2022 norma were applied for the retrospective on-site verification of method using anonymized respiratory specimens collected during the last 2024-2025 autumn-winter season in France. <b>Results:</b> Usability and satisfaction were comparable to current reference point-of-care platforms, such as the Cepheid GeneXpert<sup>®</sup> Xpress System (Cepheid Diagnostics, Sunnyvale, CA, USA). Repeatability and reproducibility (2.34-4.49% and 2.78-5.71%, respectively) demonstrated a high level of precision. The platform exhibited a low invalid rate (2.9%), with most resolving on retesting. Analytical performance on 301 clinical samples showed high overall sensitivities: 94.8% for SARS-CoV-2 (Ct ≤ 33), 95.8% for influenza A and B viruses, 95.2% for RSV, and 95.4% for all viruses. Specificities were consistently high (99.2-100.0%). False negatives (2.6%) were predominantly associated with high Ct values. Agreement with the comparator reference NeuMoDx™ Flu A-B/RSV/SARS-CoV-2 Vantage Assay (Qiagen GmbH, Hilden, Germany) was almost perfect (Cohen's κ 0.939-0.974), and a total of 91.1%, 94.8%, and 100.0% of Ct values were within the 95% limits of agreement for the detection of SARS-CoV-2, influenza A and B viruses, and RSV, respectively, by Bland-Altman analyses. Passing-Bablok regression analyses demonstrated good Ct values correlation between VitaSIRO <i>solo</i>™ and NeuMoDx™ assays, with a slight, non-significant, positive bias for the VitaSIRO <i>solo</i>™ assay (mean absolute bias +0.509 to +0.898). <b>Conclusions:</b> These findings support VitaSIRO <i>solo</i>™ Instrument as a user-friendly and reliable point-of-care platform for the rapid detection and differentiation of SARS-CoV-2, influenza A and B viruses, and RSV responding to the EN ISO 15189:2022 criteria for accreditation to be implemented in hospital or decentralized settings.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-09-05
DOI
https://doi.org/10.3390/diagnostics15172249
Citations
1
Primary Topic
Respiratory viral infections research
Type
article
Field-Weighted Citation Impact
1.35
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simultaneous Detection and Differentiation of SARS-CoV-2, Influenza A/B, and Respiratory Syncytial Viruses in Respiratory Specimens Using the VitaSIRO solo™ SARS-CoV-2/Flu/RSV Assay

Ralph‐Sydney Mboumba Bouassa, Laurent Bélec, Sarah Lukumbisa
1 citations
Diagnostics
Respiratory viral infections research
1.35
article

Simultaneous Detection and Differentiation of SARS-CoV-2, Influenza A/B, and Respiratory Syncytial Viruses in Respiratory Specimens Using the VitaSIRO solo™ SARS-CoV-2/Flu/RSV Assay

Ralph‐Sydney Mboumba Bouassa, Laurent Bélec, Sarah Lukumbisa
article en
1 citations

Abstract

<b>Background/Objectives:</b> The concurrent circulation of SARS-CoV-2 with influenza A and B viruses and respiratory syncytial virus (RSV) represents a new diagnostic challenge in the post-COVID-19 area, especially considering that these infections have overlapping clinical presentations but different approaches to treatment and management. Multiplexed molecular testing on point-of-care platforms that focus on the simultaneous detection of multiple respiratory viruses in a single tube constitutes a useful approach for diagnosis of respiratory infections in decentralized clinical settings. This study evaluated the analytical performances of the VitaSIRO <i>solo</i>™ SARS-CoV-2/Flu/RSV Assay performed on the VitaSIRO <i>solo</i>™ Instrument (Credo Diagnostics Biomedical Pte. Ltd., Singapore, Republic of Singapore). <b>Methods:</b> With a view to accreditation, the criteria of the 2022-revised EN ISO 15189:2022 norma were applied for the retrospective on-site verification of method using anonymized respiratory specimens collected during the last 2024-2025 autumn-winter season in France. <b>Results:</b> Usability and satisfaction were comparable to current reference point-of-care platforms, such as the Cepheid GeneXpert<sup>®</sup> Xpress System (Cepheid Diagnostics, Sunnyvale, CA, USA). Repeatability and reproducibility (2.34-4.49% and 2.78-5.71%, respectively) demonstrated a high level of precision. The platform exhibited a low invalid rate (2.9%), with most resolving on retesting. Analytical performance on 301 clinical samples showed high overall sensitivities: 94.8% for SARS-CoV-2 (Ct ≤ 33), 95.8% for influenza A and B viruses, 95.2% for RSV, and 95.4% for all viruses. Specificities were consistently high (99.2-100.0%). False negatives (2.6%) were predominantly associated with high Ct values. Agreement with the comparator reference NeuMoDx™ Flu A-B/RSV/SARS-CoV-2 Vantage Assay (Qiagen GmbH, Hilden, Germany) was almost perfect (Cohen's κ 0.939-0.974), and a total of 91.1%, 94.8%, and 100.0% of Ct values were within the 95% limits of agreement for the detection of SARS-CoV-2, influenza A and B viruses, and RSV, respectively, by Bland-Altman analyses. Passing-Bablok regression analyses demonstrated good Ct values correlation between VitaSIRO <i>solo</i>™ and NeuMoDx™ assays, with a slight, non-significant, positive bias for the VitaSIRO <i>solo</i>™ assay (mean absolute bias +0.509 to +0.898). <b>Conclusions:</b> These findings support VitaSIRO <i>solo</i>™ Instrument as a user-friendly and reliable point-of-care platform for the rapid detection and differentiation of SARS-CoV-2, influenza A and B viruses, and RSV responding to the EN ISO 15189:2022 criteria for accreditation to be implemented in hospital or decentralized settings.

DiagnosticsVol. 15(17)
Délégation Paris 5 (FR), University of Ottawa (CA), Université Paris-Est Créteil (FR), Université Paris Cité (FR), Montfort Hospital (CA), Assistance Publique – Hôpitaux de Paris (FR), Hôpital Européen Georges-Pompidou (FR), Hôpital Européen (FR), Centre International de Recherches Médicales de Franceville (GA), Institut du Savoir Montfort (CA)
Openalex Percentile: Top 16%
Respiratory viral infections research
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.