Enhancing blood culture volume with ultrathin-wall cannula devices

Suboptimal blood volume collection negatively affects pathogen recovery and time to pathogen detection in clinical blood culture samples. This study evaluated the performance of the ultrathin-wall cannula device of the BD Vacutainer UltraTouch Push Button Blood Collection Set (UltraTouch; BD Life Sciences, Franklin Lakes, NJ) in increasing the blood volume collected compared to regular cannulas. In this retrospective study, the volume of blood collected in negative BD BACTEC Plus Aerobic/F Culture Vials prior to and following blind conversion to UltraTouch was measured from blood background metabolic activity signal acquired during the initial incubation period in the BD BACTEC FX instrument using a custom software application (patent pending). Differences in average fill volume between pre- and post-conversion were evaluated overall and by month. A separate logistic regression was also performed to model the relationship between monthly average blood volume increase (mL) and sample positivity rate. Prevalence rates for identified organisms were also assessed. The blood fill volume of 13,356 pre-conversion and 119,971 post-conversion bottles was compared. The average post-conversion fill volume was significantly higher than that of pre-conversion (7.56 mL [95% CI, 7.54-7.59] vs. 5.68 mL [95% CI, 5.62-5.74], respectively). This 1.88 mL increase in collected blood volume increased sample positivity from 7.0% to 7.7% (<i>P</i> = 0.002; odds ratio [OR], 1.11; 95% CI, 1.04-1.19). This study demonstrated that using an ultrathin-wall cannula blood collection device can significantly increase blood culture volume and sample positivity, supporting improved identification of patients suffering from bloodstream infections.IMPORTANCESubstandard blood fill volume practices in clinical settings negatively affect pathogen recovery which, in turn, may impact patients' care. Whereas professional education regarding the importance of adhering to clinical guidelines has been helpful in partly remediating the issue, the role of blood collection set technologies in optimizing blood culture volume has not been studied as thoroughly. Our study assessed the performance of the BD Vacutainer UltraTouch Push Button Blood Collection Set, which is designed with an ultrathin-wall cannula, to improve blood culture volume.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Microbiology
Published
2025-09-05
DOI
https://doi.org/10.1128/jcm.00208-25
Primary Topic
Bacterial Identification and Susceptibility Testing
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhancing blood culture volume with ultrathin-wall cannula devices

Valentin Parvu, Grant Johnson, Arek Zubrzycki, Stéphane Beauchamp et al.
Journal of Clinical Microbiology
Bacterial Identification and Susceptibility Testing
article

Enhancing blood culture volume with ultrathin-wall cannula devices

Valentin Parvu, Grant Johnson, Arek Zubrzycki, Stéphane Beauchamp, Lauren Cooper, Mike Cuttler
article en

Abstract

Suboptimal blood volume collection negatively affects pathogen recovery and time to pathogen detection in clinical blood culture samples. This study evaluated the performance of the ultrathin-wall cannula device of the BD Vacutainer UltraTouch Push Button Blood Collection Set (UltraTouch; BD Life Sciences, Franklin Lakes, NJ) in increasing the blood volume collected compared to regular cannulas. In this retrospective study, the volume of blood collected in negative BD BACTEC Plus Aerobic/F Culture Vials prior to and following blind conversion to UltraTouch was measured from blood background metabolic activity signal acquired during the initial incubation period in the BD BACTEC FX instrument using a custom software application (patent pending). Differences in average fill volume between pre- and post-conversion were evaluated overall and by month. A separate logistic regression was also performed to model the relationship between monthly average blood volume increase (mL) and sample positivity rate. Prevalence rates for identified organisms were also assessed. The blood fill volume of 13,356 pre-conversion and 119,971 post-conversion bottles was compared. The average post-conversion fill volume was significantly higher than that of pre-conversion (7.56 mL [95% CI, 7.54-7.59] vs. 5.68 mL [95% CI, 5.62-5.74], respectively). This 1.88 mL increase in collected blood volume increased sample positivity from 7.0% to 7.7% (<i>P</i> = 0.002; odds ratio [OR], 1.11; 95% CI, 1.04-1.19). This study demonstrated that using an ultrathin-wall cannula blood collection device can significantly increase blood culture volume and sample positivity, supporting improved identification of patients suffering from bloodstream infections.IMPORTANCESubstandard blood fill volume practices in clinical settings negatively affect pathogen recovery which, in turn, may impact patients' care. Whereas professional education regarding the importance of adhering to clinical guidelines has been helpful in partly remediating the issue, the role of blood collection set technologies in optimizing blood culture volume has not been studied as thoroughly. Our study assessed the performance of the BD Vacutainer UltraTouch Push Button Blood Collection Set, which is designed with an ultrathin-wall cannula, to improve blood culture volume.

Journal of Clinical MicrobiologyVol. 63(10)
Becton Dickinson (United States) (US), Lakeridge Health (CA), Becton Dickinson (Canada) (CA)
Openalex Percentile: Top 74%
Bacterial Identification and Susceptibility Testing
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.