In radiological imaging, much research has been conducted on tube voltage and additional filters to reduce radiation exposure. In this study, the usefulness of heavy metal filters in lumbar spine imaging to maintain image quality and reduce radiation dose was investigated. A human-body phantom was irradiated with various combinations of tube voltages (70, 75, 80, 85, and 90 kV) and filters (Cu, Gd, Ho, Yb, W). For all beam qualities, the mAs value was adjusted so that the amount of CsI fluorescence at each dose after passing through the phantom was the same, and the contrast-to-noise ratio (CNR) and entrance-surface air dose (ESD) were measured. At 80 kV-Gd, the ESD was reduced by 34% while maintaining the CNR. The results showed that 80 kV-Gd gave the optimal beam quality for anterior-posterior lumbar spine imaging of patients with a standard body type.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2025-09-04
DOI
https://doi.org/10.1093/rpd/ncaf100
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimization of beam quality for lumbar spine imaging with heavy metal filters: a phantom model study

Seiichi Murakami, Chihiro Koide
Radiation Protection Dosimetry
Advanced X-ray and CT Imaging
article

Optimization of beam quality for lumbar spine imaging with heavy metal filters: a phantom model study

Seiichi Murakami, Chihiro Koide
article en

Abstract

In radiological imaging, much research has been conducted on tube voltage and additional filters to reduce radiation exposure. In this study, the usefulness of heavy metal filters in lumbar spine imaging to maintain image quality and reduce radiation dose was investigated. A human-body phantom was irradiated with various combinations of tube voltages (70, 75, 80, 85, and 90 kV) and filters (Cu, Gd, Ho, Yb, W). For all beam qualities, the mAs value was adjusted so that the amount of CsI fluorescence at each dose after passing through the phantom was the same, and the contrast-to-noise ratio (CNR) and entrance-surface air dose (ESD) were measured. At 80 kV-Gd, the ESD was reduced by 34% while maintaining the CNR. The results showed that 80 kV-Gd gave the optimal beam quality for anterior-posterior lumbar spine imaging of patients with a standard body type.

Radiation Protection DosimetryVol. 201(17)
Junshin Gakuen University (JP)
Openalex Percentile: Top 78%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.