Assessment of blood hemolysis during optimization of the RotaFlow centrifugal pump impeller

This study focuses on the evaluation of a modernized impeller for the RotaFlow centrifugal pump (Maquet, Germany), carried out as part of efforts to design a domestic counterpart. The proposed impeller features a combination of primary elongated blades, responsible for generating the majority of pressure, and secondary shortened blades. The investigation examined pump performance under extracorporeal membrane oxygenation (ECMO) therapy conditions at a pressure of 350 mmHg and flow rate of 5 L/min. Computational analyses were conducted to evaluate fluid flow parameters associated with hemolysis risk. The optimized impeller demonstrated a significant increase in low tangential stress zones (<10 Pa), reduced exposure time, and a lower hemolysis index. Comparative mathematical modeling and bench testing with donor blood confirmed the improved hemodynamic performance of the redesigned impeller over the original configuration.

Authors

Institutions

Publication Details

Journal
Russian Journal of Transplantology and Artificial Organs
Published
2025-09-03
DOI
https://doi.org/10.15825/1995-1191-2025-3-117-124
Primary Topic
Cavitation Phenomena in Pumps
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of blood hemolysis during optimization of the RotaFlow centrifugal pump impeller

A. P. Kuleshov, V. K. Bogdanov, N. V. Grudinin, А. С. Бучнев et al.
Russian Journal of Transplantology and Artificial Organs
Cavitation Phenomena in Pumps
article

Assessment of blood hemolysis during optimization of the RotaFlow centrifugal pump impeller

A. P. Kuleshov, V. K. Bogdanov, N. V. Grudinin, А. С. Бучнев, V. A. Elenkin, D. N. Shilkin
article en

Abstract

This study focuses on the evaluation of a modernized impeller for the RotaFlow centrifugal pump (Maquet, Germany), carried out as part of efforts to design a domestic counterpart. The proposed impeller features a combination of primary elongated blades, responsible for generating the majority of pressure, and secondary shortened blades. The investigation examined pump performance under extracorporeal membrane oxygenation (ECMO) therapy conditions at a pressure of 350 mmHg and flow rate of 5 L/min. Computational analyses were conducted to evaluate fluid flow parameters associated with hemolysis risk. The optimized impeller demonstrated a significant increase in low tangential stress zones (<10 Pa), reduced exposure time, and a lower hemolysis index. Comparative mathematical modeling and bench testing with donor blood confirmed the improved hemodynamic performance of the redesigned impeller over the original configuration.

Russian Journal of Transplantology and Artificial OrgansVol. 27(3)
V.I.Shumakov Federal Research Center of Transplantology and Artificial Organs (RU)
Openalex Percentile: Top 73%
Cavitation Phenomena in Pumps
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.