Assessment of blood hemolysis during optimization of the RotaFlow centrifugal pump impeller
This study focuses on the evaluation of a modernized impeller for the RotaFlow centrifugal pump (Maquet, Germany), carried out as part of efforts to design a domestic counterpart. The proposed impeller features a combination of primary elongated blades, responsible for generating the majority of pressure, and secondary shortened blades. The investigation examined pump performance under extracorporeal membrane oxygenation (ECMO) therapy conditions at a pressure of 350 mmHg and flow rate of 5 L/min. Computational analyses were conducted to evaluate fluid flow parameters associated with hemolysis risk. The optimized impeller demonstrated a significant increase in low tangential stress zones (<10 Pa), reduced exposure time, and a lower hemolysis index. Comparative mathematical modeling and bench testing with donor blood confirmed the improved hemodynamic performance of the redesigned impeller over the original configuration.
Authors
- A. P. Kuleshov
- V. K. Bogdanov (ORCID: https://orcid.org/0009-0008-8405-1716)
- N. V. Grudinin (ORCID: https://orcid.org/0000-0003-1100-9513)
- А. С. Бучнев
- V. A. Elenkin
- D. N. Shilkin
Institutions
- V.I.Shumakov Federal Research Center of Transplantology and Artificial Organs (RU)
Publication Details
- Journal
- Russian Journal of Transplantology and Artificial Organs
- Published
- 2025-09-03
- DOI
- https://doi.org/10.15825/1995-1191-2025-3-117-124
- Primary Topic
- Cavitation Phenomena in Pumps
- Type
- article
- Field-Weighted Citation Impact
- 0.00