Development of a Novel Endotracheal Tube to Measure Posterior Laryngeal Pressure
ABSTRACT Objectives The development of an endotracheal tube (ETT) prototype to measure the pressure on the posterior glottis during intubation. Methods Three ETTs , size 6, 7, and 8 mm, were modified with a piezoelectric pressure sensor. These ETTs were then used to intubate a laryngeal model. Pressures were measured at a hypopharyngeal laryngeal angle ( HLA ) of 90°, and the laryngeal platform was raised to create a more acute HLA ranging from 90° to 45°. These same ETTs were also used to intubate the larynx of a fresh frozen cadaver, with pressure measured with the head in neutral and flexed positions. Results The pressures exerted by the 6 mm ETT ranged from 7.33 cm H 2 O at 90 degrees HLA to 62.57 cm H 2 O at 45 degrees HLA . The 7 mm ETT ranged from 5.42 to 451.19 cm H 2 O and the 8 mm ETT pressure ranged from 136.98 to 534.38 cm H 2 O , respectively. The differences were statistically significant. The cadaveric study revealed the average pressure for the 6‐, 7‐, and 8‐mm ETT to be 2.03, 166.78, and 228.49 cm H 2 O , respectively, in the neutral position and 0.85, 30.37, and 158.51 cm H 2 O , respectively, in the flexed position. These differences were statistically significant. Conclusion Posterior laryngeal pressure monitoring can be conducted utilizing ETT‐based pressure sensors. Progression of this technology may help minimize laryngeal injury caused by intubation by determining which patients may be optimal candidates for early tracheostomy. Level of Evidence 5
Authors
- Denis Lafreniere (ORCID: https://orcid.org/0000-0002-4021-7576)
- Amelia Geist
Institutions
- UConn Health (US)
Publication Details
- Journal
- The Laryngoscope
- Published
- 2025-09-04
- DOI
- https://doi.org/10.1002/lary.70113
- Primary Topic
- Airway Management and Intubation Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- University of Connecticut