Comparative in vitro evaluation of trueness and precision in five intraoral scanners across multiple clinical scenarios

With rapid technological advancements, numerous intraoral scanners (IOS) are available. This study evaluates and compares the trueness and precision of five IOS across four clinical scenarios to provide benchmark comparative data. Resin models included a completely edentulous maxilla, a dentulous maxilla, a lower jaw with 10 prepared teeth, and an edentulous mandible with six implant scanbodies. Reference scans were obtained using a high-resolution extraoral scanner. Each model was scanned 10 times with five IOS (Aoralscan 3, CS 3600, iTero Element 5D, TRIOS3, Virtuo Vivo). 3D stereolithography files were created, and IOS scans were compared to reference scans using reverse engineering software. NURBS/NURBS scanbody library file comparison was performed on implant models. Statistical analysis (SPSS 26) assessed trueness and precision (root mean square values), using ANOVA and post-hoc tests (p < 0.05). TRIOS3 had the best trueness in all groups except group D (iTero: 22.6 ± 2.4 μm). TRIOS3 also showed the best precision across all groups (e.g., group D: 14.5 ± 7.6 μm). Significant differences were found between IOS and between groups using the same IOS (p < 0.05). ICC tests indicated excellent correlation, and Dahlberg’s formula confirmed minimal method error. Within the limitations of this in vitro study, TRIOS3 and iTero Element 5D demonstrated higher trueness and precision among the tested devices. These findings should be regarded as reference benchmarks, and future studies should validate them in vivo while assessing next-generation IOS. This study provides methodological benchmark data on older-generation IOS models. While TRIOS3 and iTero Element 5D performed better than other devices tested, clinical translation requires caution, and newer scanners must be evaluated in future studies. • Study comparing five IOS across dentate, edentulous, prepared teeth, and implant clinical scenarios. • TRIOS3 showed superior trueness & precision in most clinical scenarios. • iTero Element 5D outperformed other scanners in dentate models. • Edentulous scans had highest errors due to lack of reference structures. • NURBS/NURBS comparison validated TRIOS3’s accuracy for implant prosthetics.

Authors

Institutions

Publication Details

Journal
Digital Dentistry Journal
Published
2025-09-04
DOI
https://doi.org/10.1016/j.ddj.2025.100039
Primary Topic
Dental materials and restorations
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative in vitro evaluation of trueness and precision in five intraoral scanners across multiple clinical scenarios

Gennaro Ruggiero, Fernando Zarone, Pietro Venezia, Roberto Sorrentino et al.
Digital Dentistry Journal
Dental materials and restorations
article

Comparative in vitro evaluation of trueness and precision in five intraoral scanners across multiple clinical scenarios

Gennaro Ruggiero, Fernando Zarone, Pietro Venezia, Roberto Sorrentino, Matteo Cariotti
article en

Abstract

With rapid technological advancements, numerous intraoral scanners (IOS) are available. This study evaluates and compares the trueness and precision of five IOS across four clinical scenarios to provide benchmark comparative data. Resin models included a completely edentulous maxilla, a dentulous maxilla, a lower jaw with 10 prepared teeth, and an edentulous mandible with six implant scanbodies. Reference scans were obtained using a high-resolution extraoral scanner. Each model was scanned 10 times with five IOS (Aoralscan 3, CS 3600, iTero Element 5D, TRIOS3, Virtuo Vivo). 3D stereolithography files were created, and IOS scans were compared to reference scans using reverse engineering software. NURBS/NURBS scanbody library file comparison was performed on implant models. Statistical analysis (SPSS 26) assessed trueness and precision (root mean square values), using ANOVA and post-hoc tests (p < 0.05). TRIOS3 had the best trueness in all groups except group D (iTero: 22.6 ± 2.4 μm). TRIOS3 also showed the best precision across all groups (e.g., group D: 14.5 ± 7.6 μm). Significant differences were found between IOS and between groups using the same IOS (p < 0.05). ICC tests indicated excellent correlation, and Dahlberg’s formula confirmed minimal method error. Within the limitations of this in vitro study, TRIOS3 and iTero Element 5D demonstrated higher trueness and precision among the tested devices. These findings should be regarded as reference benchmarks, and future studies should validate them in vivo while assessing next-generation IOS. This study provides methodological benchmark data on older-generation IOS models. While TRIOS3 and iTero Element 5D performed better than other devices tested, clinical translation requires caution, and newer scanners must be evaluated in future studies. • Study comparing five IOS across dentate, edentulous, prepared teeth, and implant clinical scenarios. • TRIOS3 showed superior trueness & precision in most clinical scenarios. • iTero Element 5D outperformed other scanners in dentate models. • Edentulous scans had highest errors due to lack of reference structures. • NURBS/NURBS comparison validated TRIOS3’s accuracy for implant prosthetics.

Digital Dentistry JournalVol. 2(2)
University of Catania (IT), Federico II University Hospital (IT), University of Naples Federico II (IT)
Openalex Percentile: Top 75%
Dental materials and restorations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.