Integration of capillary chromatography columns into a 3D-printed microfluidic multiport valve for miniaturized multi-column chromatography

Driven by the need for smaller and more efficient continuous chromatographic systems, this study explores the feasibility of miniaturizing multi-column chromatography by integrating monolithic capillary columns directly into a 3D-printed valve rotor. First, monolith capillary ion-exchange columns were synthesized in PEEK capillaries with a diameter of 0.75 mm and a 3D-printed rotor system that can hold these capillary columns was developed. Single-column experiments confirm that the synthesized monolithic columns can separate the proteins hemoglobin and lysozyme with step elution, although the dynamic binding capacity is significantly lower than that of commercial equivalents. Subsequent multi-column tests reveal that a standard gradient simulated moving bed (SMB) approach faces limitations due to elevated back pressure of the columns, resulting in leakages. To overcome this issue, a continuous multi-column chromatography (CMCC) setup was implemented, enabling promising separations with more than 60% yield and 80% purity for both proteins with switching intervals as low as 30s, allowing for a fast response time of the system. Although further improvements in monolith morphology, functionalization, and valve sealing are needed, these findings highlight the potential of integrating monolith columns in miniaturized multi-column processes.

Authors

Institutions

Publication Details

Journal
Journal of Chromatography A
Published
2025-08-15
DOI
https://doi.org/10.1016/j.chroma.2025.466245
Primary Topic
Analytical Chemistry and Chromatography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Integration of capillary chromatography columns into a 3D-printed microfluidic multiport valve for miniaturized multi-column chromatography

David Achauer, Juliane Diehm, Matthias Franzreb
Journal of Chromatography A
Analytical Chemistry and Chromatography
article

Integration of capillary chromatography columns into a 3D-printed microfluidic multiport valve for miniaturized multi-column chromatography

David Achauer, Juliane Diehm, Matthias Franzreb
article en

Abstract

Driven by the need for smaller and more efficient continuous chromatographic systems, this study explores the feasibility of miniaturizing multi-column chromatography by integrating monolithic capillary columns directly into a 3D-printed valve rotor. First, monolith capillary ion-exchange columns were synthesized in PEEK capillaries with a diameter of 0.75 mm and a 3D-printed rotor system that can hold these capillary columns was developed. Single-column experiments confirm that the synthesized monolithic columns can separate the proteins hemoglobin and lysozyme with step elution, although the dynamic binding capacity is significantly lower than that of commercial equivalents. Subsequent multi-column tests reveal that a standard gradient simulated moving bed (SMB) approach faces limitations due to elevated back pressure of the columns, resulting in leakages. To overcome this issue, a continuous multi-column chromatography (CMCC) setup was implemented, enabling promising separations with more than 60% yield and 80% purity for both proteins with switching intervals as low as 30s, allowing for a fast response time of the system. Although further improvements in monolith morphology, functionalization, and valve sealing are needed, these findings highlight the potential of integrating monolith columns in miniaturized multi-column processes.

Journal of Chromatography AVol. 1760
Karlsruhe Institute of Technology (DE)
Openalex Percentile: Top 79%
Analytical Chemistry and Chromatography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Integration of capillary chromatography columns into a 3D-printed microfluidic multiport valve for miniaturized multi-column chromatography — David Achauer, Juliane Diehm, et al. · Journal of Chromatography A (2025) | TGRS Research Map | TGRS