Estimation of mean glandular dose using breast-equivalent phantoms: a foundational study for regional DRL formulation in mammography in India

The objective of this foundational study is to support the development of regional diagnostic reference levels (DRLs) for mammography screening in India by estimating mean glandular dose (MGD) using appropriate breast-equivalent phantoms across computed mammography (CR), digital mammography (DR), and digital breast tomosynthesis (DBT) systems. Additionally, system-displayed MGD values were compared with calculated MGD values to evaluate their accuracy for routine clinical use. A total of 5000 mammographic views were collected across CR, DR, and DBT units. For each unit and phantom thickness, ten entrance surface air kerma (ESAK) measurements were performed under clinical settings using a calibrated solid-state dosimeter (AccuGold+, Radcal, USA). Breast-equivalent slab phantoms (Sun Nuclear, USA), ranging from 30 mm to 90 mm, were used to simulate varying glandular compositions. Standard American College of Radiology and DBT phantoms were employed to ensure consistent imaging conditions. MGDs were derived from ESAK values using IAEA recommended conversion factors. For DR and DBT systems, calculated MGDs were compared with system displayed values. The 75th percentile MGDs for each phantom thickness were used to propose regional DRLs, in line with ICRP Publication 135 and IAEA Human Health Report No. 77. At the standard 60 mm compressed breast thickness, the 75th percentile MGDs were found to be 2.58 ± 0.38 mGy (CR), 1.68 ± 0.86 mGy (DR), and 2.66 ± 0.78 mGy (DBT). System-displayed MGDs in DR and DBT were observed to be 13%-17% higher than calculated values. This study presents a scalable, reproducible framework for establishing regional, thickness-specific DRLs in India, supporting dose optimisation and quality assurance in mammography practice.

Authors

Institutions

Publication Details

Journal
Journal of Radiological Protection
Published
2025-09-01
DOI
https://doi.org/10.1088/1361-6498/ae02a4
Citations
1
Primary Topic
Digital Radiography and Breast Imaging
Type
article
Field-Weighted Citation Impact
1.52

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Estimation of mean glandular dose using breast-equivalent phantoms: a foundational study for regional DRL formulation in mammography in India

Balvinder Kaur Sapra, Saravana Kumar A, KN Govindarajan, S. Prasanna et al.
1 citations
Journal of Radiological Protection
Digital Radiography and Breast Imaging
1.52
article

Estimation of mean glandular dose using breast-equivalent phantoms: a foundational study for regional DRL formulation in mammography in India

Balvinder Kaur Sapra, Saravana Kumar A, KN Govindarajan, S. Prasanna, Sudhir Kumar, Thangavel Manojkumar
article en
1 citations

Abstract

The objective of this foundational study is to support the development of regional diagnostic reference levels (DRLs) for mammography screening in India by estimating mean glandular dose (MGD) using appropriate breast-equivalent phantoms across computed mammography (CR), digital mammography (DR), and digital breast tomosynthesis (DBT) systems. Additionally, system-displayed MGD values were compared with calculated MGD values to evaluate their accuracy for routine clinical use. A total of 5000 mammographic views were collected across CR, DR, and DBT units. For each unit and phantom thickness, ten entrance surface air kerma (ESAK) measurements were performed under clinical settings using a calibrated solid-state dosimeter (AccuGold+, Radcal, USA). Breast-equivalent slab phantoms (Sun Nuclear, USA), ranging from 30 mm to 90 mm, were used to simulate varying glandular compositions. Standard American College of Radiology and DBT phantoms were employed to ensure consistent imaging conditions. MGDs were derived from ESAK values using IAEA recommended conversion factors. For DR and DBT systems, calculated MGDs were compared with system displayed values. The 75th percentile MGDs for each phantom thickness were used to propose regional DRLs, in line with ICRP Publication 135 and IAEA Human Health Report No. 77. At the standard 60 mm compressed breast thickness, the 75th percentile MGDs were found to be 2.58 ± 0.38 mGy (CR), 1.68 ± 0.86 mGy (DR), and 2.66 ± 0.78 mGy (DBT). System-displayed MGDs in DR and DBT were observed to be 13%-17% higher than calculated values. This study presents a scalable, reproducible framework for establishing regional, thickness-specific DRLs in India, supporting dose optimisation and quality assurance in mammography practice.

Journal of Radiological ProtectionVol. 45(3)
Bhabha Atomic Research Centre (IN), PSG Institute of Medical Sciences & Research (IN), Homi Bhabha National Institute (IN), Bhabha Atomic Research Center Hospital (IN), PSG INSTITUTE OF TECHNOLOGY AND APPLIED RESEARCH (IN)
Board of Research in Nuclear Sciences
Openalex Percentile: Top 15%
Digital Radiography and Breast Imaging
1.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.