2.5D imaging: obtaining depth information from 2D helium-beam radiographs<sup>*</sup>

<i>Objective.</i>Ion-beam radiography has been proposed as a daily on-couch imaging modality for ion-beam radiotherapy range verification. However, an ion-beam radiograph only contains 2D information since it shows a projection of the patient along the beam direction. To extract depth information of anatomical changes from 2D helium-beam radiographs, we experimentally investigated the 2.5D imaging approach.<i>Approach.</i>At the Heidelberg ion-beam therapy center, we acquired helium-beam radiographs of a homogeneous plastic phantom and an anthropomorphic phantom. We changed the geometry of both phantoms to mimic anatomical changes and acquired data before and after the change. After reconstructing plane-of-interest difference images along depth in 1 mm steps, we analyzed the spatial resolution by means of the slanted edge approach and a method based on a median filter and Tenengrad focus measure. The depth of the maximum spatial resolution was then compared to the one where the change was introduced in the experiment and their difference was determined as measure of accuracy.<i>Main results.</i>For both the slanted edge and Tenengrad method, all accuracy values of the plastic phantom were below 11 mm. Furthermore, the found depth values of all but one gap position coincided with the ground truth position of the edge within the 2 <i>σ</i>uncertainty. In the anthropomorphic phantom, we reached a depth accuracy of13-7+5 mm.<i>Significance.</i>The 2.5D imaging approach, which was investigated experimentally for the first time in this work, offers a technique to estimate the depth of an anatomical change based on the acquisition of 2D ion-beam radiographs. This information is crucial to determine whether the change is located within the treatment field and thus impairs the treatment. 2.5D ion-beam radiography could therefore be used for daily on-couch imaging to monitor the anatomy of the patient and make ion-beam radiotherapy treatments more precise.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-09-03
DOI
https://doi.org/10.1088/1361-6560/ae02de
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

2.5D imaging: obtaining depth information from 2D helium-beam radiographs*

Annika Schlechter, Margareta Metzner, Tim Gehrke, Daria Zhevachevska et al.
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
article

2.5D imaging: obtaining depth information from 2D helium-beam radiographs*

Annika Schlechter, Margareta Metzner, Tim Gehrke, Daria Zhevachevska, O. Jaekel, P. G. Schlegel, M. Martišíková
article en

Abstract

<i>Objective.</i>Ion-beam radiography has been proposed as a daily on-couch imaging modality for ion-beam radiotherapy range verification. However, an ion-beam radiograph only contains 2D information since it shows a projection of the patient along the beam direction. To extract depth information of anatomical changes from 2D helium-beam radiographs, we experimentally investigated the 2.5D imaging approach.<i>Approach.</i>At the Heidelberg ion-beam therapy center, we acquired helium-beam radiographs of a homogeneous plastic phantom and an anthropomorphic phantom. We changed the geometry of both phantoms to mimic anatomical changes and acquired data before and after the change. After reconstructing plane-of-interest difference images along depth in 1 mm steps, we analyzed the spatial resolution by means of the slanted edge approach and a method based on a median filter and Tenengrad focus measure. The depth of the maximum spatial resolution was then compared to the one where the change was introduced in the experiment and their difference was determined as measure of accuracy.<i>Main results.</i>For both the slanted edge and Tenengrad method, all accuracy values of the plastic phantom were below 11 mm. Furthermore, the found depth values of all but one gap position coincided with the ground truth position of the edge within the 2 <i>σ</i>uncertainty. In the anthropomorphic phantom, we reached a depth accuracy of13-7+5 mm.<i>Significance.</i>The 2.5D imaging approach, which was investigated experimentally for the first time in this work, offers a technique to estimate the depth of an anatomical change based on the acquisition of 2D ion-beam radiographs. This information is crucial to determine whether the change is located within the treatment field and thus impairs the treatment. 2.5D ion-beam radiography could therefore be used for daily on-couch imaging to monitor the anatomy of the patient and make ion-beam radiotherapy treatments more precise.

Physics in Medicine and BiologyVol. 70(20)
German Cancer Research Center (DE), Heidelberg University (DE), University Hospital Heidelberg (DE), Heidelberger Institut für Radioonkologie (DE), National Center for Tumor Diseases (DE), University Medical Centre Mannheim (DE)
Deutsche Forschungsgemeinschaft
Openalex Percentile: Top 66%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.