A58 Radiation dose reduction with ultra-low dose photon-counting CT for adult ventriculoperitoneal shunt assessment compared to conventional X-ray

<h3>Introduction</h3> Conventional X-rays have so far been used for the radiological monitoring of ventriculoperitoneal (VP) shunts. The cumulative radiation dose is not negligible. Photon counting computed tomography enables good image quality even with low-dose protocols. <h3>Aim of study</h3> The aim of this study was to compare the radiation doses of ultra-low dose PCCT with conventional radiographic VP shunt series in the assessment of shunt pathways in adults. <h3>Method</h3> This retrospective study included 15 ultra-low dose non-contrast PCCT scans in adults (covering the VP shunt pathway from the skull or temporal bone to the ischium) and 52 radiographic VP shunt (each with a.p. skull, a.p. chest, a.p. and lateral abdomen). The effective doses from both imaging modalities were compared using the Mann-Whitney U test. <h3>Results</h3> Ultra-low dose PCCT had a significantly lower radiation dose in terms of effective dose compared to the radiographic shunt series, with an estimated median effective dose of 0.12 mSv (IQR 0.10–0.17 mSv, including the topogram) compared to the estimated median total effective dose for the radiographic VP shunt series of 0.31 mSv (IQR 0.22–0.46 mSv) (p&lt;0.001). This results in a 60% median radiation dose reduction with PCCT compared to conventional shunt series. <h3>Conclusion</h3> Ultra-low dose PCCT provides a significant reduction in radiation dose of approximately 60% compared to conventional X-ray series for assessing VP shunts in adults, with the added benefit of superior visualisation. However, availability and cost must also be considered. <h3>Conflict of Interest</h3> Yes J.R. Kroeger received research support from Philips, support for attending meetings and/or travel from Veryan, honoraria for scientific lectures from GE Healthcare and honoraria for clinical advisory board membership from Siemens Healthineers. J. Borggrefe received honoraria for scientific lectures from Siemens Healthineers and Philips Healthcare.

Authors

Institutions

Publication Details

Published
2025-09-01
DOI
https://doi.org/10.1136/jnis-2025-esmint.41
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A58 Radiation dose reduction with ultra-low dose photon-counting CT for adult ventriculoperitoneal shunt assessment compared to conventional X-ray

Christoph Moenninghoff, Sebastian Horstmeier, Jan Borggrefe, Cornelius Deuschl et al.
Radiation Dose and Imaging
article

A58 Radiation dose reduction with ultra-low dose photon-counting CT for adult ventriculoperitoneal shunt assessment compared to conventional X-ray

Christoph Moenninghoff, Sebastian Horstmeier, Jan Borggrefe, Cornelius Deuschl, Iram Shahzadi, Arwed Elias Michael, Denise Bos, Ali Al-Omari, D Schönbeck, Ulrich J. Knappe, Michael Forsting, Jan Robert Kröger
article en

Abstract

<h3>Introduction</h3> Conventional X-rays have so far been used for the radiological monitoring of ventriculoperitoneal (VP) shunts. The cumulative radiation dose is not negligible. Photon counting computed tomography enables good image quality even with low-dose protocols. <h3>Aim of study</h3> The aim of this study was to compare the radiation doses of ultra-low dose PCCT with conventional radiographic VP shunt series in the assessment of shunt pathways in adults. <h3>Method</h3> This retrospective study included 15 ultra-low dose non-contrast PCCT scans in adults (covering the VP shunt pathway from the skull or temporal bone to the ischium) and 52 radiographic VP shunt (each with a.p. skull, a.p. chest, a.p. and lateral abdomen). The effective doses from both imaging modalities were compared using the Mann-Whitney U test. <h3>Results</h3> Ultra-low dose PCCT had a significantly lower radiation dose in terms of effective dose compared to the radiographic shunt series, with an estimated median effective dose of 0.12 mSv (IQR 0.10–0.17 mSv, including the topogram) compared to the estimated median total effective dose for the radiographic VP shunt series of 0.31 mSv (IQR 0.22–0.46 mSv) (p&lt;0.001). This results in a 60% median radiation dose reduction with PCCT compared to conventional shunt series. <h3>Conclusion</h3> Ultra-low dose PCCT provides a significant reduction in radiation dose of approximately 60% compared to conventional X-ray series for assessing VP shunts in adults, with the added benefit of superior visualisation. However, availability and cost must also be considered. <h3>Conflict of Interest</h3> Yes J.R. Kroeger received research support from Philips, support for attending meetings and/or travel from Veryan, honoraria for scientific lectures from GE Healthcare and honoraria for clinical advisory board membership from Siemens Healthineers. J. Borggrefe received honoraria for scientific lectures from Siemens Healthineers and Philips Healthcare.

University Hospitals of the Ruhr-University of Bochum (DE), Essen University Hospital (DE), Siemens Healthineers (Germany) (DE), Johannes Wesling Klinikum Minden (DE), University of Duisburg-Essen (DE)
Openalex Percentile: Top 67%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.