Reprocessing of single-use energy devices: efficacy cleaning aspect

Objectives: Energy devices (EDs), such as Harmonic, Ligasure, Thunderbeat and Trocar are widely used in Minimally Invasive Surgeries. They are expensive and designed for single use. However, due to the limitation of resources, they have been reused in some cases. Therefore, we aimed to assess the efficacy of EDs reprocessing by Adenosine Triphosphate (ATP) method. Methods: This was a cross-sectional description study. After first clinical using, EDs were taken to cleaning. Every ED was cleaned three times. Efficacy cleaning was assessed after each cleaning procedure by ATP method. ATP <200 RLUs (Relative Light Units) was benchmark as efficient cleaning process. Results: A total of 611 EDs were studied, including 269 of Harmonic, Ligasure, Thunderbeat, 298 of Trocar, and 44 other types. Detachable devices accounted for about 32.7%. Overall, after three consecutive cleanings, the median ATP values were decreased dramatically (957 RLUs, 160 RLUs, and 62 RLUs, respectively). This was a significant reduction in ATP levels between three stages (p < 0.05). There were 63.5%, 84.3%, and 92.8% EDs that had ATP < 200 RLUs after first, second, third cleaning respectively. Approximately 90% of EDs were still functional after three cleaning times. Nondetachable items were to be more difficult to clean than detachable ones (p = 0.0003, OR 1.3 [1.1 – 1.5]). Conclusions: Our data suggest that monitoring efficacy cleaning of surgical instruments in general and single-use energy devices in particular with ATP can identify a number of different influence factors, like the instrument condition, reprocessing procedure, or especially their structure. ATP measurement seems to be a valid technique that allows an immediate repeat of the manual cleaning if the results exceed the established cutoff of 200 RLUs.

Authors

Institutions

Publication Details

Journal
Antimicrobial Stewardship & Healthcare Epidemiology
Published
2025-05-01
DOI
https://doi.org/10.1017/ash.2025.78
Primary Topic
Recycling and Waste Management Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reprocessing of single-use energy devices: efficacy cleaning aspect

Thanh Tran, Duy H. N. Nguyen, Huỳnh Minh Tuấn, Haiyan Lu et al.
Antimicrobial Stewardship & Healthcare Epidemiology
Recycling and Waste Management Techniques
article

Reprocessing of single-use energy devices: efficacy cleaning aspect

Thanh Tran, Duy H. N. Nguyen, Huỳnh Minh Tuấn, Haiyan Lu, Thoa Trinh, Nguyen Thi Hai Yen, Tran Thien Dat Nguyen, Thuy T. Pham, Huyen Nguyen, Thi Kim Ngan Duong, Văn Nam Thái, C. T. Vu
article en

Abstract

Objectives: Energy devices (EDs), such as Harmonic, Ligasure, Thunderbeat and Trocar are widely used in Minimally Invasive Surgeries. They are expensive and designed for single use. However, due to the limitation of resources, they have been reused in some cases. Therefore, we aimed to assess the efficacy of EDs reprocessing by Adenosine Triphosphate (ATP) method. Methods: This was a cross-sectional description study. After first clinical using, EDs were taken to cleaning. Every ED was cleaned three times. Efficacy cleaning was assessed after each cleaning procedure by ATP method. ATP <200 RLUs (Relative Light Units) was benchmark as efficient cleaning process. Results: A total of 611 EDs were studied, including 269 of Harmonic, Ligasure, Thunderbeat, 298 of Trocar, and 44 other types. Detachable devices accounted for about 32.7%. Overall, after three consecutive cleanings, the median ATP values were decreased dramatically (957 RLUs, 160 RLUs, and 62 RLUs, respectively). This was a significant reduction in ATP levels between three stages (p < 0.05). There were 63.5%, 84.3%, and 92.8% EDs that had ATP < 200 RLUs after first, second, third cleaning respectively. Approximately 90% of EDs were still functional after three cleaning times. Nondetachable items were to be more difficult to clean than detachable ones (p = 0.0003, OR 1.3 [1.1 – 1.5]). Conclusions: Our data suggest that monitoring efficacy cleaning of surgical instruments in general and single-use energy devices in particular with ATP can identify a number of different influence factors, like the instrument condition, reprocessing procedure, or especially their structure. ATP measurement seems to be a valid technique that allows an immediate repeat of the manual cleaning if the results exceed the established cutoff of 200 RLUs.

Antimicrobial Stewardship & Healthcare EpidemiologyVol. 5(S1)
University of Medicine and Pharmacy at Ho Chi Minh City (VN), University Medical Center (US)
Openalex Percentile: Top 71%
Recycling and Waste Management Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.