Evaluation of loop-mediated isothermal amplification (LAMP) assay for detection of five bacterial periprosthetic joint infection

Objectives: Periprosthetic joint infection (PJI) is one of the most serious and debilitating complications that can occur after total joint arthroplasty. Therefore, early diagnosis and appropriate treatment are important for a good prognosis. Recently, molecular diagnostic methods have been widely used to detect the causative microorganisms of PJI sensitively and rapidly. The Multiplex Loop- Mediated Isothermal Amplification (LAMP) method is faster and easier to perform compared to polymerase chain reaction (PCR)-based assays. Therefore, this study developed a multiplex LAMP assay for diagnosing bacterial PJI using LAMP technology and evaluated its analytical and clinical performance. Methods: We developed a multiplex LAMP assay for the detection of five bacteria: Staphylococcus aureus, Staphylococcus epidermidis, Streptococcus agalactiae, Pseudomonas aeruginosa, and Escherichia coli, frequently observed to be the causative agents of PJI. The method limit of detection (LOD) and cross-reactivity were determined by spiking standard strains into the joint synovial fluid. The LOD of the multiplex LAMP assay was compared with that of a quantitative real-time PCR (qPCR) assay. Clinical performance was evaluated using 20 joint synovial fluid samples collected from patients suspected of having bacterial PJI. Results: The LOD of the gram-positive bacterial multiplex LAMP assay and qPCR were 10 5 /10 4 CFU/mL, 10 3 /10 3 CFU/mL, and 10 5 /10 4 CFU/mL against S. agalactiae , S. epidermidis, and S. aureus , respectively. For P. aeruginosa and E. coli , the LOD of the multiplex LAMP and qPCR assays were 10 5 /10 4 and 10 6 /10 4 CFU/mL, respectively. The multiplex LAMP assay detects target bacteria without cross-reacting with other bacteria, and exhibited 100% sensitivity and specificity in clinical performance evaluation. Conclusions: This multiplex LAMP assay can rapidly detect five high-prevalence bacterial species causing bacterial PJI, with excellent sensitivity and specificity, in less than 1 h, and it may be useful for the early diagnosis of PJI.

Authors

Institutions

Publication Details

Journal
Antimicrobial Stewardship & Healthcare Epidemiology
Published
2025-05-01
DOI
https://doi.org/10.1017/ash.2025.158
Primary Topic
Orthopedic Infections and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of loop-mediated isothermal amplification (LAMP) assay for detection of five bacterial periprosthetic joint infection

Ji‐Hoon Bae, Chae Seung Lim, Woong Sik Jang, Min Cho
Antimicrobial Stewardship & Healthcare Epidemiology
Orthopedic Infections and Treatments
article

Evaluation of loop-mediated isothermal amplification (LAMP) assay for detection of five bacterial periprosthetic joint infection

Ji‐Hoon Bae, Chae Seung Lim, Woong Sik Jang, Min Cho
article en

Abstract

Objectives: Periprosthetic joint infection (PJI) is one of the most serious and debilitating complications that can occur after total joint arthroplasty. Therefore, early diagnosis and appropriate treatment are important for a good prognosis. Recently, molecular diagnostic methods have been widely used to detect the causative microorganisms of PJI sensitively and rapidly. The Multiplex Loop- Mediated Isothermal Amplification (LAMP) method is faster and easier to perform compared to polymerase chain reaction (PCR)-based assays. Therefore, this study developed a multiplex LAMP assay for diagnosing bacterial PJI using LAMP technology and evaluated its analytical and clinical performance. Methods: We developed a multiplex LAMP assay for the detection of five bacteria: Staphylococcus aureus, Staphylococcus epidermidis, Streptococcus agalactiae, Pseudomonas aeruginosa, and Escherichia coli, frequently observed to be the causative agents of PJI. The method limit of detection (LOD) and cross-reactivity were determined by spiking standard strains into the joint synovial fluid. The LOD of the multiplex LAMP assay was compared with that of a quantitative real-time PCR (qPCR) assay. Clinical performance was evaluated using 20 joint synovial fluid samples collected from patients suspected of having bacterial PJI. Results: The LOD of the gram-positive bacterial multiplex LAMP assay and qPCR were 10 5 /10 4 CFU/mL, 10 3 /10 3 CFU/mL, and 10 5 /10 4 CFU/mL against S. agalactiae , S. epidermidis, and S. aureus , respectively. For P. aeruginosa and E. coli , the LOD of the multiplex LAMP and qPCR assays were 10 5 /10 4 and 10 6 /10 4 CFU/mL, respectively. The multiplex LAMP assay detects target bacteria without cross-reacting with other bacteria, and exhibited 100% sensitivity and specificity in clinical performance evaluation. Conclusions: This multiplex LAMP assay can rapidly detect five high-prevalence bacterial species causing bacterial PJI, with excellent sensitivity and specificity, in less than 1 h, and it may be useful for the early diagnosis of PJI.

Antimicrobial Stewardship & Healthcare EpidemiologyVol. 5(S1)
Korea University (KR)
Openalex Percentile: Top 62%
Orthopedic Infections and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.